弥合化学发光免疫测定和液相色谱-串联质谱监测英夫利昔单抗之间的差距:方法验证和临床可比性研究
Bridging the Gap Between Chemiluminescence Immunoassay and Liquid Chromatography-Tandem Mass Spectrometry for Infliximab Monitoring: Method Validation and Clinical Comparability Study.
Bridging the Gap Between Chemiluminescence Immunoassay and Liquid Chromatography-Tandem Mass Spectrometry for Infliximab Monitoring: Method Validation and Clinical Comparability Study.
英夫利西单抗(IFX)是儿童炎症性肠病治疗的基石生物制剂,治疗药物监测(TDM)至关重要,但受平台间差异影响而复杂化。因此,有必要对自动化免疫分析法与参考方法进行系统比较。本研究旨在验证一种用于IFX定量检测的室内化学发光免疫分析法(CLIA),并评估其与液相色谱-串联质谱法(LC-MS/MS)的一致性。通过评估空白限、特异性、线性、准确度、精密度、基质等效性和稳定性对CLIA试剂盒进行验证。采用CLIA和参考LC-MS/MS方法测定52份儿童血浆样本中的IFX浓度。使用线性回归、Passing-Bablok回归和Bland-Altman分析评估可比性。CLIA表现出优异的性能(定量下限:0.640 µg/mL;特异性干扰<10.0%;可报告范围:0.640-462 µg/mL)。LC-MS/MS测得的浓度高于CLIA(中位数:6.25 vs. 2.87 µg/mL)。两者呈强相关性(r = 0.9101),线性无显著偏差(Passing-Bablok,p > 0.05)。然而,Bland-Altman分析显示在治疗范围内(3-7 µg/mL)存在显著的平均相对偏倚(67.9%)和较差的分级一致性(52%符合率),表明两者不可互换。我们验证了一种适用于高通量TDM的快速自动化CLIA。尽管相关性较强,但显著的定量差异排除了直接结果替换的可能性,凸显了建立方法特异性治疗阈值的必要性。
Infliximab (IFX) is a cornerstone biologic for pediatric inflammatory bowel disease, where therapeutic drug monitoring (TDM) is essential but complicated by inter-platform variability. Systematic comparison of automated immunoassays with reference methods is therefore warranted. This study aimed to validate an in-house chemiluminescence immunoassay (CLIA) for IFX quantification and evaluate its agreement with liquid chromatography-tandem mass spectrometry (LC-MS/MS). A CLIA kit was validated by assessing the limit of blank, specificity, linearity, accuracy, precision, matrix equivalence, and stability. IFX concentrations in 52 pediatric plasma samples were measured by CLIA and a reference LC-MS/MS method. Comparability was evaluated using linear regression, Passing-Bablok regression, and Bland-Altman analysis. CLIA showed excellent performance (lower limit of quantification: 0.640 µg/mL; specificity <10.0% interference; reportable range: 0.640-462 µg/mL). LC-MS/MS yielded higher concentrations than the CLIA (median: 6.25 vs. 2.87 µg/mL). Strong correlation was observed (r = 0.9101) with no significant deviation from linearity (Passing-Bablok, p > 0.05). However, Bland-Altman revealed substantial mean relative bias (67.9%) and poor categorical agreement (52% concordance) within the therapeutic range (3-7 µg/mL), indicating non-interchangeability. We validated a rapid, automated CLIA suitable for high-throughput TDM. Despite strong correlation, significant quantitative differences preclude direct result substitution, underscoring the need for method-specific therapeutic thresholds.
Ying Xia, Ming-Mei Zhu, Meng-Qing Xiao, Chen-Chun Zhong, Yu-Jun Zhou, Yi-Hao Xue, et al. Bridging the Gap Between Chemiluminescence Immunoassay and Liquid Chromatography-Tandem Mass Spectrometry for Infliximab Monitoring: Method Validation and Clinical Comparability Study.. Journal of separation science. 2026; doi:10.1002/jssc.70515.
Ying Xia, Ming-Mei Zhu, Meng-Qing Xiao, Chen-Chun Zhong, Yu-Jun Zhou, Yi-Hao Xue, et al (2026). Bridging the Gap Between Chemiluminescence Immunoassay and Liquid Chromatography-Tandem Mass Spectrometry for Infliximab Monitoring: Method Validation and Clinical Comparability Study.. Journal of separation science. https://doi.org/10.1002/jssc.70515
Ying Xia, Ming-Mei Zhu, Meng-Qing Xiao, Chen-Chun Zhong, Yu-Jun Zhou, Yi-Hao Xue, et al. Bridging the Gap Between Chemiluminescence Immunoassay and Liquid Chromatography-Tandem Mass Spectrometry for Infliximab Monitoring: Method Validation and Clinical Comparability Study.[J]. Journal of separation science, 2026 doi:10.1002/jssc.70515.