新型双能射频/脉冲场消融导管的安全性和有效性:FlexPulse研究的12个月结果

Safety and Effectiveness of a Novel Dual-Energy Radiofrequency/Pulsed Field Ablation Catheter: 12 Month Results of the FlexPulse Study.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

脉冲电场消融越来越多地用于治疗阵发性心房颤动。FlexPulse研究旨在评估一种新型双能量、柔性尖端聚焦消融导管(具备射频和脉冲电场功能)治疗有症状、药物难治性阵发性心房颤动的安全性和有效性。FlexPulse是一项全球性、前瞻性、单臂、多中心研究。消融手术以肺静脉隔离为目标,并由术者酌情进行辅助消融。主要安全性终点为任何消融手术后7天内发生预先定义的安全性事件。主要有效性终点为12个月内无复发性房颤/房性心动过速/心房扑动、重复手术、新使用或升级I/III类抗心律失常药物和/或心脏电复律的复合终点。共有180名受试者接受了阵发性心房颤动消融。所有参与者均成功实现肺静脉隔离。37.2%(67/180)的受试者接受了额外的非肺静脉消融。主要安全性终点事件发生率为1.7%。12个月时主要有效性终点为74.6%,无记录的房颤/心房扑动/房性心动过速复发率为77.4%(在不进行每周经电话监测的情况下为92.6%)。在空白期后接受重复消融的15名受试者中,再次标测显示79.7%(47/59)的肺静脉在指数手术后仍保持持久隔离。回顾性分析发现,采用脉冲电场指数≥30且AutoMark距离目标达6 mm的工作流程,其12个月无房颤/心房扑动/房性心动过速率为84.6%。在本试验中,使用TactiFlex Duo对有症状、药物难治性阵发性心房颤动患者进行消融是安全且有效的。

Abstract

Pulsed field (PF) ablation is increasingly used to treat paroxysmal atrial fibrillation (PAF). FlexPulse (NCT06676072) was conducted to evaluate safety and effectiveness of a novel dual-energy, flexible-tip focal ablation catheter with radiofrequency and PF for treatment of symptomatic, drug-refractory PAF. FlexPulse was a global, prospective, single arm, multicentre study. The ablation procedures targeted pulmonary vein isolation (PVI) and adjunctive ablation at operators' discretion. The primary safety endpoint was occurrence of pre-defined safety events within 7-days of any ablation procedure. The primary effectiveness endpoint was freedom from a composite of recurrent AF/atrial tachycardia/atrial flutter, repeat procedure, new use or escalation of class I/III AAD, and/or cardioversion through 12-months. Overall, 180 subjects underwent PAF ablation. Successful PVI was achieved in all participants. Additional non-PV ablations were performed in 37.2% (67/180) of subjects. The primary safety endpoint event rate was 1.7%. The primary effectiveness endpoint was 74.6% at 12-months, and freedom from documented AF/AFL/AT recurrence was 77.4% (92.6% without weekly transtelephonic monitoring). In 15 subjects who underwent repeat ablation after the blanking period, remapping indicated that 79.7% (47/59) of the PVs were durably isolated from the index procedures. Retrospective analysis found that workflows with PFA Index (PI) ≥30 and AutoMark distance up to 6 mm target had a 12-month freedom from AF/AFL/AT of 84.6%. In this trial, ablation with TactiFlex Duo in patients with symptomatic, drug-refractory PAF was safe and effective.

如何引用

AMA

Ayman A Hussein, Gediminas Rackauskas, Sri Sundaram, Jonathan P Piccini, Franz Xaver Roithinger, Patricia Tung, et al. Safety and Effectiveness of a Novel Dual-Energy Radiofrequency/Pulsed Field Ablation Catheter: 12 Month Results of the FlexPulse Study.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2026; doi:10.1093/europace/euag238.

APA

Ayman A Hussein, Gediminas Rackauskas, Sri Sundaram, Jonathan P Piccini, Franz Xaver Roithinger, Patricia Tung, et al (2026). Safety and Effectiveness of a Novel Dual-Energy Radiofrequency/Pulsed Field Ablation Catheter: 12 Month Results of the FlexPulse Study.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. https://doi.org/10.1093/europace/euag238

GB/T 7714

Ayman A Hussein, Gediminas Rackauskas, Sri Sundaram, Jonathan P Piccini, Franz Xaver Roithinger, Patricia Tung, et al. Safety and Effectiveness of a Novel Dual-Energy Radiofrequency/Pulsed Field Ablation Catheter: 12 Month Results of the FlexPulse Study.[J]. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology, 2026 doi:10.1093/europace/euag238.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI

Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology 的其他文献