半生理盐水灌注与生理盐水灌注在增加开放式灌注温控消融电流输送方面的益处有限

Limited benefit of half-normal vs normal saline irrigation for augmenting current delivery with open-irrigated temperature-controlled ablation.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

半正常盐水(HNS)灌注通过增强电流向组织的传递,在功率控制消融过程中增大射频(RF)损伤范围。这一优势在基于导管尖端温度调节功率输出的温度控制消融中是否依然存在,目前尚不清楚。本研究旨在离体模型中比较使用HNS与正常盐水(NS)灌注时,功率控制与温度控制开放灌注导管的损伤特征。使用功率控制和温度控制灌注导管,分别以NS和HNS灌注对猪左心室组织施加射频消融。消融在10 g接触力下垂直于组织进行。温度控制消融采用制造商预设的最高温度截止值,即30W时为50°C,50W时为55°C。在分析的104处损伤中,功率控制导管使用HNS灌注在30W×30秒时产生的损伤大于NS(深度:7.5 ± 1.0 vs 6.5 ± 0.6 mm,= .0005;宽度:10.5 ± 1.1 vs 9.4 ± 1.0 mm,= .005)。然而,温度控制导管使用HNS与NS在30W时(深度:6.1 ± 0.4 vs 6.2 ± 0.7 mm,= .44;宽度:9.4 ± 0.7 vs 9.6 ± 1.2 mm,= .23)或50W时(深度:7.2 ± 0.7 vs 7.5 ± 0.8 mm,= .35;宽度:10.9 ± 1.3 vs 11.8 ± 1.6 mm,= .19)均未观察到损伤尺寸差异。使用HNS时,温度控制导管较NS显示出平均功率输出降低(30W:23.3 ± 5.4 vs 27.1 ± 4.6 W,< .05)且灌注流速增加,这可能归因于温度截止值的执行。使用HNS时,温度控制导管观察到的蒸汽爆裂少于功率控制导管(5% vs 23%,= .09)。与功率控制消融不同,HNS灌注在温度控制消融中不增大损伤尺寸。最高温度截止值的执行降低了实际功率输出,削弱了HNS灌注的电流增强优势,同时可能提高安全性。

Abstract

Half-normal saline (HNS) irrigation increases radiofrequency (RF) lesion size during power-controlled ablation by augmenting current delivery to tissue. Whether this benefit persists with temperature-controlled ablation, which modulates power delivery on the basis of catheter tip temperature, remains unknown. This study sought to compare lesion characteristics using HNS vs normal saline (NS) irrigation with power-controlled and temperature-controlled open-irrigated catheters in an ex vivo model. RF applications were delivered to left ventricular porcine tissue using a power- and temperature-controlled irrigated catheter with NS and HNS irrigation. Lesions were performed at 10 g contact force perpendicular to tissue. Temperature-controlled ablation used manufacturer-preset maximum temperature cutoffs of 50°C at 30W and 55°C at 50W. Among 104 lesions analyzed, HNS irrigation with the power-controlled catheter produced larger lesions at 30W × 30 seconds than did NS (depth: 7.5 ± 1.0 vs 6.5 ± 0.6 mm, = .0005; width: 10.5 ± 1.1 vs 9.4 ± 1.0 mm, = .005). However, no difference in lesion dimensions was observed with the temperature-controlled catheter using HNS vs NS at 30W (depth: 6.1 ± 0.4 vs 6.2 ± 0.7 mm, = .44; width: 9.4 ± 0.7 vs 9.6 ± 1.2 mm, = .23) or 50W (depth: 7.2 ± 0.7 vs 7.5 ± 0.8 mm, = .35; width: 10.9 ±1.3 vs 11.8 ± 1.6 mm, = .19). With HNS, the temperature-controlled catheter showed reduced average power delivery (30W: 23.3 ± 5.4 vs 27.1± 4.6 W, < .05) and increased irrigation flow rates compared with NS, likely owing to temperature cut-off enforcement. Fewer steam pops were observed with the temperature-controlled than with the power-controlled catheter using HNS (5% vs 23%, = .09). Unlike power-controlled ablation, HNS irrigation does not increase lesion size with temperature-controlled ablation. Enforcement of maximum temperature cutoffs reduces actual power delivery, blunting the current augmentation benefit of HNS irrigation while potentially improving safety.

如何引用

AMA

Henry D Huang, Jonathan Gordon, Jeffrey Winterfield, Pasquale Santangeli, Jackson Liang, Hagai Yavin, et al. Limited benefit of half-normal vs normal saline irrigation for augmenting current delivery with open-irrigated temperature-controlled ablation.. Heart rhythm O2. 2026; doi:10.1016/j.hroo.2026.05.006.

APA

Henry D Huang, Jonathan Gordon, Jeffrey Winterfield, Pasquale Santangeli, Jackson Liang, Hagai Yavin, et al (2026). Limited benefit of half-normal vs normal saline irrigation for augmenting current delivery with open-irrigated temperature-controlled ablation.. Heart rhythm O2. https://doi.org/10.1016/j.hroo.2026.05.006

GB/T 7714

Henry D Huang, Jonathan Gordon, Jeffrey Winterfield, Pasquale Santangeli, Jackson Liang, Hagai Yavin, et al. Limited benefit of half-normal vs normal saline irrigation for augmenting current delivery with open-irrigated temperature-controlled ablation.[J]. Heart rhythm O2, 2026 doi:10.1016/j.hroo.2026.05.006.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI

Heart rhythm O2 的其他文献