脉冲电场消融(PFA)存在溶血风险;然而,3种单次放电PFA系统之间的差异仍不明确。本研究旨在探讨3种PFA系统在溶血方面的差异。我们评估了131例接受3种PFA系统治疗的心房颤动患者(VARIPULSE,n=26;PulseSelect,n=62;FARAPULSE,n=43)在基线及消融后24小时溶血、心肌和肾脏生物标志物的变化,同时评估了23例接受射频消融的患者(对照组)和8例未接受消融的患者(未消融对照组)。各组基线特征具有可比性。结合珠蛋白水平在VARIPULSE组从97.3±52.1降至49.9±40.1,在PulseSelect组从96.7±41.9降至53.5±31.2,在FARAPULSE组从90.6±43.8降至27.3±18.6(均P<.001)。与对照组相比,PFA组的溶血和心肌损伤均更为明显,其中FARAPULSE组严重程度最高。消融后各组均未观察到肾功能显著恶化;然而,FARAPULSE组发生2例1期急性肾损伤,两例均经水化治疗后完全缓解。在所有PFA组中,消融后结合珠蛋白水平与放电次数呈显著负相关。多因素回归分析确定放电比率(总放电次数/制造商推荐放电次数)和体重指数是结合珠蛋白水平降低的独立预测因素。血管内溶血是PFA后值得关注的问题,与放电次数和体重指数相关,且3种PFA系统之间可能存在差异。
Pulsed field ablation (PFA) is associated with a risk of hemolysis; however, the differences among 3 single-shot PFA systems remain unclear. This study aimed to examine the differences in hemolysis among 3 PFA systems. We assessed changes in hemolytic, myocardial, and renal biomarkers at baseline and 24 hours after ablation in 131 patients with atrial fibrillation treated with 3 PFA systems (VARIPULSE, n = 26; PulseSelect, n = 62; FARAPULSE, n = 43), as well as in 23 patients undergoing radiofrequency ablation (control) and 8 patients without ablation (nonablation control). Baseline characteristics were comparable across the groups. Haptoglobin levels decreased from 97.3 ± 52.1 to 49.9 ± 40.1 in the VARIPULSE group, 96.7 ± 41.9 to 53.5 ± 31.2 in the PulseSelect group, and 90.6 ± 43.8 to 27.3 ± 18.6 in the FARAPULSE group (all < .001). Compared with the control group, both hemolysis and myocardial injury were more pronounced in the PFA groups, with the greatest severity observed in the FARAPULSE group. No significant deterioration in renal function was observed in any group after ablation; however, 2 cases of stage 1 acute kidney injury occurred in the FARAPULSE group, both of which fully resolved with hydration. Postablation haptoglobin levels exhibited a significant negative correlation with the number of applications across all PFA groups. Multivariate regression analysis identified the application ratio (total number of applications/manufacturer's recommended applications) and body mass index as independent predictors of reduced haptoglobin levels. Intravascular hemolysis is a pertinent concern after PFA and is associated with the number of applications and body mass index, with potential differences across the 3 PFA systems.
Hiroyuki Miyazawa, Satoshi Yanagisawa, Yasuya Inden, Taishi Fukushima, Takehiro Hiramatsu, Kentaro Adachi, et al. Comparative analysis of hemolysis incidence across 3 different pulsed field ablation systems.. Heart rhythm O2. 2026; doi:10.1016/j.hroo.2026.04.009.
Hiroyuki Miyazawa, Satoshi Yanagisawa, Yasuya Inden, Taishi Fukushima, Takehiro Hiramatsu, Kentaro Adachi, et al (2026). Comparative analysis of hemolysis incidence across 3 different pulsed field ablation systems.. Heart rhythm O2. https://doi.org/10.1016/j.hroo.2026.04.009
Hiroyuki Miyazawa, Satoshi Yanagisawa, Yasuya Inden, Taishi Fukushima, Takehiro Hiramatsu, Kentaro Adachi, et al. Comparative analysis of hemolysis incidence across 3 different pulsed field ablation systems.[J]. Heart rhythm O2, 2026 doi:10.1016/j.hroo.2026.04.009.
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出