利用锥形几何和二维交叉扫描OCT实现快速的机器人到iOCT校准

Fast robot-to-iOCT calibration using conical geometry and 2D cross-scan OCT.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

光学相干断层扫描(OCT)与机器人坐标系之间的精确空间配准对于OCT引导的显微手术至关重要,其中精确的工具定位对于安全、可靠的介入操作必不可少。我们提出并验证了一种在机器人显微手术系统中校准OCT与机器人坐标系之间变换的实用且准确的方法。将一个刚性连接于机器人腕部的锥形校准工具使用快速二维OCT横截面进行扫描。经过去噪、边缘提取和双线拟合后,对锥尖进行定位,并将所得坐标纳入基于向量的三点校准中以恢复OCT到机器人的变换。该方法支持手动和自动操作,包括调整位姿以改善扫描平面 alignment。使用扫频源OCT系统、Meca500机器人和3D打印校准工具进行的实验表明,在OCT工作范围内配准稳定。该方法在存在OCT噪声和锥体几何形状中等制造缺陷的情况下也保持稳健。单次锥尖定位的重复性估计为 ,其中在16 mm的OCT X-Y横向视场范围内进行校准,在台式环境中转化为0.31%的线性校准不确定度。所提出的基于锥体的框架为高精度手术系统中常规的机器人到OCT校准提供了一种实用且准确的解决方案。

Abstract

Accurate spatial registration between optical coherence tomography (OCT) and robotic coordinate systems is critical for OCT-guided microsurgery, where precise tool localization is essential for safe, reliable intervention. We propose and validate a practical and accurate method for calibrating the transformation between OCT and robot coordinate systems in robotic microsurgical systems. A conical calibration tool rigidly attached to the robot wrist was scanned using fast 2D OCT cross-sections. After denoising, edge extraction, and two-line fitting, the cone tip was localized, and the resulting coordinates were incorporated into a vector-based three-point calibration to recover the OCT-to-robot transformation. The method supports both manual and automated operation, including pose adjustment to improve scan-plane alignment. Experiments with a swept-source OCT system, a Meca500 robot, and a 3D-printed calibration tool showed stable registration across the OCT working range. The method also remained robust in the presence of OCT noise and moderate manufacturing imperfections of the cone geometry. The single cone-tip localization repeatability was estimated to be , where calibrating over 16 mm of OCT X-Y lateral field-of-view range translates to 0.31% of linear calibration uncertainty in a bench-top environment. The proposed cone-based framework provides a practical and accurate solution for routine robot-to-OCT calibration in high-precision surgical systems.

如何引用

AMA

Krzysztof Gromada, Tomasz Piesio, Piotr Kasprzycki, Andrea Curatolo, Karol Karnowski. Fast robot-to-iOCT calibration using conical geometry and 2D cross-scan OCT.. Journal of biomedical optics. 2026; doi:10.1117/1.JBO.31.12.123305.

APA

Krzysztof Gromada, Tomasz Piesio, Piotr Kasprzycki, Andrea Curatolo, Karol Karnowski (2026). Fast robot-to-iOCT calibration using conical geometry and 2D cross-scan OCT.. Journal of biomedical optics. https://doi.org/10.1117/1.JBO.31.12.123305

GB/T 7714

Krzysztof Gromada, Tomasz Piesio, Piotr Kasprzycki, Andrea Curatolo, Karol Karnowski. Fast robot-to-iOCT calibration using conical geometry and 2D cross-scan OCT.[J]. Journal of biomedical optics, 2026 doi:10.1117/1.JBO.31.12.123305.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI

Journal of biomedical optics 的其他文献