模拟咳嗽方案下老年合成疝网的生物力学行为

Biomechanical Behavior of an Aged Synthetic Hernia Mesh Under a Cough-Simulating Protocol.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

疝补片在日常活动中会反复承受由腹腔内压力波动引起的生理性载荷。本研究旨在探讨模拟反复剧烈咳嗽的长时间循环加载对过期、未植入的Microval聚丙烯补片力学性能的影响。研究在两种主要材料方向上对老化Microval补片试样进行了单轴循环拉伸试验。加载方案模拟反复咳嗽,连续三天每天施加六次加载。在整个实验过程中评估残余应变、弹性模量和加载持续时间。加载方案结束后,对试样进行至破坏的测试。循环加载对补片产生了影响。在第三个测试日之后,纵向(L)方向的应变在12-20%以内,横向(T)方向高达25%。弹性模量随反复加载显著增加,表明材料明显硬化,而极限抗拉强度仍处于未使用试样的测量范围内。此外,第一个加载周期比后续周期长约20%。长时间循环加载导致老化Microval补片产生显著的永久变形和明显的材料硬化,而其极限抗拉强度仅受到轻微影响。这些发现增进了对疝补片在反复生理性加载下力学行为的理解。

Abstract

Hernia meshes are repeatedly subjected to physiological loading caused by fluctuations in intra-abdominal pressure during daily activities. This study aimed to investigate the effects of prolonged cyclic loading simulating repetitive severe coughing on the mechanical properties of expired, non-implanted Microval polypropylene mesh. Uniaxial cyclic tensile tests were performed on aged Microval mesh specimens in the two principal material directions. The loading protocol simulated repetitive coughing by applying six loading sessions per day over three consecutive days. Residual strain, elastic modulus, and loading duration were evaluated throughout the experiment. After the loading protocol, the specimens were tested to failure. Cyclic loading affected the mesh. Strain was within 12-20% in the longitudinal (L) direction and up to 25% in the transverse (T) direction after the third testing day. The elastic modulus increased significantly with repetitive loading, indicating pronounced material stiffening, while the ultimate tensile strength remained within the range measured for the unused specimens. In addition, the first loading cycle was approximately 20% longer than the subsequent cycles. Prolonged cyclic loading caused significant permanent deformation and pronounced material stiffening in the aged Microval mesh, whereas its ultimate tensile strength was only slightly affected. These findings enhance understanding of the mechanical behavior of hernia meshes under repetitive physiological loading.

如何引用

AMA

Miglena G Kirilova-Doneva, Dessislava L Pashkouleva. Biomechanical Behavior of an Aged Synthetic Hernia Mesh Under a Cough-Simulating Protocol.. Materials (Basel, Switzerland). 2026; doi:10.3390/ma19163368.

APA

Miglena G Kirilova-Doneva, Dessislava L Pashkouleva (2026). Biomechanical Behavior of an Aged Synthetic Hernia Mesh Under a Cough-Simulating Protocol.. Materials (Basel, Switzerland). https://doi.org/10.3390/ma19163368

GB/T 7714

Miglena G Kirilova-Doneva, Dessislava L Pashkouleva. Biomechanical Behavior of an Aged Synthetic Hernia Mesh Under a Cough-Simulating Protocol.[J]. Materials (Basel, Switzerland), 2026 doi:10.3390/ma19163368.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI

Materials (Basel, Switzerland) 的其他文献