老年患者髋部骨折仍是重大临床挑战,最佳手术策略尚存争议。关于采用延长偏心距骨水泥柄行全髋关节置换术(THR)治疗老年股骨颈骨折(FNF)后外展肌力量恢复及功能结局的证据仍然有限。我们评估了老年FNF患者行THR的功能结局和外展肌力量,比较了延长内侧偏心距(MO)柄与标准MO柄。我们回顾性分析了120例采用延长MO Exeter 44 mm柄行THR的FNF患者(A组)和100例采用标准MO Exeter 37.5 mm柄行THR的FNF患者(B组)。纳入患者Koval分级为3级或以上、美国麻醉医师协会(ASA)分级为3级或以下、简易精神状态检查评分24分或以上。比较患侧髋关节术后MO与自体髋关节。使用肌力测力计(KINVENT technologies)测量外展肌力量。调查包括脱位、大腿疼痛和股骨大转子疼痛(GTP)在内的并发症。两组间ASA、体重指数、骨密度或假体位置均无显著差异。A组患侧髋关节平均MO为42.6 mm,B组为36.2 mm,A组较自体髋关节(37.8 mm)大约5 mm,B组较自体髋关节(30.4 mm)大约6 mm。A组患侧髋关节外展肌力量恢复至自体髋关节的约86%,B组为81%。未发生脱位或大腿疼痛。A组有5例患者出现轻度GTP,接受物理治疗。与B组相比,A组患侧与自体髋关节在平均MO、平均垂直偏心距和外展肌力量方面的差异更小(< 0.05)。对于接受THR的健康老年FNF患者,延长MO柄对外展肌力量恢复具有积极作用,可能有助于改善关节稳定性。
Hip fractures in elderly patients remain a major clinical challenge, with optimal surgical strategies still debated. Evidence regarding abductor strength recovery and functional outcomes after total hip replacement (THR) using an extended offset cemented stem in elderly patients with femoral neck fracture (FNF) remains limited. We evaluated functional outcomes and abductor muscle strength of THR for FNF in elderly patients, comparing an extended medial offset (MO) stem and a standard MO stem. We retrospectively reviewed 120 patients undergoing THR with an extended MO Exeter 44-mm stem (group A) and 100 patients undergoing THR with a standard MO Exeter 37.5-mm stem (group B) for FNF. We enrolled patients with a Koval grade of 3 or higher, an American Society of Anesthesiologists (ASA) grade of 3 or lower, and a Mini-Mental State Examination score of 24 or higher. Postoperative MO of the affected hip was compared with the native hip. Abductor muscle strength was measured using a muscle dynamometer (KINVENT technologies). Complications including dislocation, thigh pain, and greater trochanteric pain (GTP) were investigated. No significant differences in ASA, body mass index, bone mineral density, or implant position were observed between the 2 groups. The mean MO of the affected hip was 42.6 mm in group A and 36.2 mm in group B, approximately 5 mm greater than the native hip in group A (37.8 mm) and 6 mm greater in group B (30.4 mm). Abductor muscle power of the affected hip recovered to approximately 86% of the native hip in group A and 81% in group B. No dislocations or thigh pain occurred. In group A, 5 patients had mild GTP treated with physical therapy. Group A demonstrated smaller differences between the affected and native hips in mean MO, mean vertical offset, and abductor muscle strength compared with group B ( < 0.05). For healthy elderly patients with FNF undergoing THR, the extended MO stem had a positive effect on the recovery of abductor muscle strength, potentially contributing to improved joint stability.
Min Seok Kim, Suc Hyun Kweon, Han Seok Ko, Jong Seok Baik. Does an Extended Offset Stem Improve Hip Stability and Function in Elderly Patients with a Femoral Neck Fracture?. Clinics in orthopedic surgery. 2026; doi:10.4055/cios25266.
Min Seok Kim, Suc Hyun Kweon, Han Seok Ko, Jong Seok Baik (2026). Does an Extended Offset Stem Improve Hip Stability and Function in Elderly Patients with a Femoral Neck Fracture?. Clinics in orthopedic surgery. https://doi.org/10.4055/cios25266
Min Seok Kim, Suc Hyun Kweon, Han Seok Ko, Jong Seok Baik. Does an Extended Offset Stem Improve Hip Stability and Function in Elderly Patients with a Femoral Neck Fracture?[J]. Clinics in orthopedic surgery, 2026 doi:10.4055/cios25266.