模型生物(如线虫)的生物相容性水凝胶封装,相较于化学或机械固定方法具有显著优势,后者可能损害生物健康并妨碍长期成像研究。本研究介绍了一种自动化、温和的方法,通过流动光刻和机器学习辅助视觉检测,实现水凝胶块内的连续固定。为优化这一自动化线虫固定过程,我们考察了流速、溶液组成和线虫加载浓度等工作参数。在人工智能线虫检测达到90%精确度的条件下,我们在低至中等线虫加载浓度(每微升1-7条线虫)下实现了86%的封装效率。较高的加载浓度(每微升10-15条线虫)使封装效率降至71%。线虫加载浓度、组成以及两种输入溶液——线虫悬浮水溶液(WSAS)和聚乙二醇二丙烯酸酯(PEGDA)溶液(PEGS)——的流速比,显著影响了线虫通量、封装过程中的线虫脱水以及封装后线虫的存活率。WSAS与PEGS的流速比设定为0.5,WSAS最低流速需达到1 μL/min以避免回流。PEGS中PEGDA浓度保持在50%以下以避免线虫脱水。通过改变水凝胶前体PEGDA的浓度,我们分别实现了适用于短期研究和长期研究的部分固定和完全固定。结果表明,封装后的线虫在3天内保持存活,凸显了该方法在不损害线虫健康或运动的情况下进行长期研究的潜力。

Abstract

Biocompatible hydrogel encapsulation of model organisms, such as , is highly desirable over chemical or mechanical immobilization methods that can harm health and hinder long term imaging studies. This study introduces an automated, gentle method for continuous immobilization within hydrogel blocks, achieved through flow lithography and machine learning-assisted vision detection. To optimize this automated worm immobilization, we investigated operating parameters, such as flow rate, solution composition, and worm loading concentration. With 90% precision in AI worm detection, we achieved 86% encapsulation efficiency at low to moderate worm-loading concentrations (1-7 worms per μL). Higher loading concentrations of 10-15 worms per μL reduced the encapsulation efficiency to 71%. Worm loading concentration, composition, and flow-rate ratios of the two input solutions-the worm-suspended aqueous solution (WSAS) and the polyethylene glycol diacrylate (PEGDA) solution (PEGS)-significantly affected worm throughput, worm dehydration during encapsulation, and the viability of encapsulated worms. The WSAS-to-PEGS flow-rate ratio was set at 0.5, with a minimum WSAS flow rate of 1 μL min required to avoid backflow. The PEGDA concentration in PEGS was kept below 50% to avoid worm dehydration. By varying the concentrations of hydrogel precursor, PEGDA, we enabled both partial and complete immobilization suitable for short term and long-term studies, respectively. Results show the encapsulated worm's viability over 3 days, highlighting this method's potential for long-term studies without compromising worm health or movement.

如何引用

AMA

Sinthuran Jegatheeswaran, Teo Mesrkhani, Wei Liu, Mohammad Kashif, June H Tan, Basel A Khader, et al. {"_":"On-demand microfluidic encapsulation of in hydrogel for long-term observation AI detection.","i":["Caenorhabditis elegans","via"]}. Lab on a chip. 2026; doi:10.1039/d6lc00148c.

APA

Sinthuran Jegatheeswaran, Teo Mesrkhani, Wei Liu, Mohammad Kashif, June H Tan, Basel A Khader, et al (2026). {"_":"On-demand microfluidic encapsulation of in hydrogel for long-term observation AI detection.","i":["Caenorhabditis elegans","via"]}. Lab on a chip. https://doi.org/10.1039/d6lc00148c

GB/T 7714

Sinthuran Jegatheeswaran, Teo Mesrkhani, Wei Liu, Mohammad Kashif, June H Tan, Basel A Khader, et al. {"_":"On-demand microfluidic encapsulation of in hydrogel for long-term observation AI detection.","i":["Caenorhabditis elegans","via"]}[J]. Lab on a chip, 2026 doi:10.1039/d6lc00148c.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI

Lab on a chip 的其他文献