Novel technique for leadless pacemaker implantation in the small right atrial appendage: a case report.
Novel technique for leadless pacemaker implantation in the small right atrial appendage: a case report.
心房无导线起搏器(Aveir AR [Abbott])近期已可供临床使用,能够实现心房起搏并维持严格的心房心室同步。尽管无导线起搏器的有效性和安全性已有报道,但部分病例因解剖变异而在植入过程中面临挑战。一名66岁男性因头晕就诊于我院。12导联心电图显示交界性心律,心率40次/分。由于患者每日口服类固醇且有囊袋感染风险,遂植入无导线起搏器而非经静脉起搏器。通过偏转操作将Aveir AR输送导管引导至右心耳(RAA)基底部。然而,在释放偏转时,由于右心耳显著较小,导管尖端滑向上腔静脉。为避免导管尖端滑脱,在保持偏转的同时将导管杆顺时针旋转,使V形标记从初始位置旋转四分之一。在保持导管偏转状态下进行的标测测量结果良好。即使在释放偏转后,导管尖端位置仍稳定于右心耳基底部而无滑脱。随后,释放输送导管偏转,并在透视下将V形标记旋转1 1/2。Aveir AR位置及各项最终参数均良好。未记录到包括心脏压塞和脱位在内的不良事件。本病例突出了在右心耳显著较小的情况下植入Aveir AR时采用保持偏转下部分旋转的特定补救技术;然而,该技术不应取代标准植入技术。
An atrial leadless pacemaker (Aveir AR [Abbott]) has recently become available, enabling atrial pacing and maintaining strict atrioventricular synchronization. Although the efficacy and safety of leadless pacemakers have already been reported, some cases present challenges during implantation due to anatomical variations. A 66-year-old man visited our hospital complaining of dizziness. A 12-lead electrocardiogram exhibited junctional rhythm with a heart rate of 40 beats/min. A leadless pacemaker was implanted instead of a transvenous pacemaker because of daily oral steroid administration and the risk of a pocket infection.The Aveir AR delivery catheter was guided to the base of the right atrial appendage (RAA) using deflection manoeuvres. However, upon deflection release, the catheter tip slipped towards the superior vena cava because of the significantly small RAA. To avoid catheter tip slip, the catheter lever was turned clockwise to rotate the chevron marker one-quarter from its initial position while maintaining deflection. Mapping measurements obtained while maintaining catheter deflection were favourable. Even after releasing the deflection, the catheter tip position remained stable at the RAA base without slippage. Subsequently, the delivery catheter deflection was released, and the chevron marker 1 1/2 was rotated under fluoroscopy. The Aveir AR position and each final parameter were favourable. No adverse events, including cardiac tamponade and dislodgement, were recorded. This case highlights the specific bailout technique using partial rotation while maintaining deflection for Aveir AR implantation in a markedly small RAA; however, it should not replace standard implantation technique.
Hirotaka Watanabe, Yoshiaki Mizutani, Junji Watanabe, Satoshi Yanagisawa, Toyoaki Murohara. Novel technique for leadless pacemaker implantation in the small right atrial appendage: a case report.. European heart journal. Case reports. 2026; doi:10.1093/ehjcr/ytag514.
Hirotaka Watanabe, Yoshiaki Mizutani, Junji Watanabe, Satoshi Yanagisawa, Toyoaki Murohara (2026). Novel technique for leadless pacemaker implantation in the small right atrial appendage: a case report.. European heart journal. Case reports. https://doi.org/10.1093/ehjcr/ytag514
Hirotaka Watanabe, Yoshiaki Mizutani, Junji Watanabe, Satoshi Yanagisawa, Toyoaki Murohara. Novel technique for leadless pacemaker implantation in the small right atrial appendage: a case report.[J]. European heart journal. Case reports, 2026 doi:10.1093/ehjcr/ytag514.
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出