FARAPULSETM脉冲场消融过程中J型尖端导丝卡在肺静脉内:病例报告
{"_":"Entrapment of a J-tip guidewire within the pulmonary vein during FARAPULSE pulsed field ablation: a case report.","sup":["TM"]}
{"_":"Entrapment of a J-tip guidewire within the pulmonary vein during FARAPULSE pulsed field ablation: a case report.","sup":["TM"]}
脉冲场消融(PFA)是一种快速应用于心房颤动(AF)消融的技术,具有良好的安全性特征和较低的并发症风险。然而,机械性并发症鲜有报道。一名37岁亚洲男性阵发性AF患者使用FARAPULSE系统接受PFA。在成功消融左上肺静脉(PV)后,J头导丝进入左下肺静脉并完全嵌顿。静脉造影显示,尽管导丝尖端仍位于血管内,但近端节段似乎沿血管壁走行,提示其穿过了管腔内结构。多次采用推拉手法尝试撤出均未成功。使用射频消融导管重新定位也未能解除嵌顿。鉴于血管破裂和大出血的风险,手术在杂交手术室转为全身麻醉,并备外科待命。桥接封堵球囊辅助取出成功解除了嵌顿导丝,未造成血管损伤。取出的导丝附着有白色纤维组织,提示其被膜性肺静脉结构嵌顿。球囊辅助血管内取出可能是处理此类嵌顿的一种安全且创伤较小的替代外科干预的方法。术者应了解这一罕见但可能危险的并发症。
Pulsed field ablation (PFA) is a rapidly adopted technology for atrial fibrillation (AF) ablation and is associated with a favourable safety profile and a low risk of complications. However, mechanical complications have rarely been reported. A 37-year-old Asian male with paroxysmal AF underwent PFA using the FARAPULSE system. Following successful ablation of the left superior pulmonary vein (PV), the J-tip guidewire advanced into the left inferior PV and became completely entrapped. Venography demonstrated that, although the guidewire tip remained intravascular, a proximal segment appeared to course along the vessel wall, suggesting passage through an intraluminal structure. Multiple attempts at withdrawal using push-pull manoeuvres were unsuccessful. Repositioning with a radiofrequency ablation catheter also failed to release the entrapment. Given the risk of vessel rupture and major haemorrhage, the procedure was converted to general anaesthesia in a hybrid operating room with surgical standby. Bridge occlusion balloon-assisted extraction successfully released the entrapped guidewire without vascular injury. The retrieved guidewire had adherent fibrous white tissue, suggesting entrapment by a membranous pulmonary venous structure. Balloon-assisted endovascular retrieval may represent a safe and less invasive alternative to surgical intervention for managing such entrapments. Operators should be aware of this rare but potentially hazardous complication.
Ken Watanabe, Takanori Arimoto, Yuta Kobayashi, Naoaki Hashimoto, Masafumi Watanabe. {"_":"Entrapment of a J-tip guidewire within the pulmonary vein during FARAPULSE pulsed field ablation: a case report.","sup":["TM"]}. European heart journal. Case reports. 2026; doi:10.1093/ehjcr/ytag503.
Ken Watanabe, Takanori Arimoto, Yuta Kobayashi, Naoaki Hashimoto, Masafumi Watanabe (2026). {"_":"Entrapment of a J-tip guidewire within the pulmonary vein during FARAPULSE pulsed field ablation: a case report.","sup":["TM"]}. European heart journal. Case reports. https://doi.org/10.1093/ehjcr/ytag503