不确定条件下离心式血液泵的鲁棒优化与随机仿真

Robust optimization and stochastic simulation of a centrifugal blood pump under uncertain conditions.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

本文旨在通过降低血泵对运行不确定性的敏感性来实现稳健优化。此外,研究了物理与运行不确定性对稳健优化后离心血泵的水动力学及血液相容性特性的影响。基线设计为一款离心血泵,其灵感来源于用作左心室辅助装置的商用泵。数值分析采用SST [公式:见原文] 湍流模型及溶血幂律模型。物理与运行条件被视为不确定的,并采用Beta概率分布函数描述。为量化不确定性,采用非侵入式多项式混沌方法,并利用Sobol'指数评估各随机参数对关注量的影响,即敏感性分析。本研究的主要目标是在保持泵工作点的同时,最小化泵效率与扬程系数的变化,确保稳健优化设计的最大溶血指数不超过基线设计。为此,通过混合进化算法进行稳健优化。优化结果清楚表明,稳健优化设计对物理与运行不确定性的敏感性较低。此外,敏感性分析与叶轮机械理论相符。泵效率和最大溶血指数主要受血液黏度变化的影响,而扬程系数主要受转速和流量的影响。泵的速度场在扩压器区域受质量流量影响较大,在其他区域受转速影响较大。

Abstract

The current paper aims to robustly optimize a blood pump by reducing its sensitivity to operational uncertainties. Moreover, the effects of physical and operational uncertainties on the hydrodynamic and hemocompatibility characteristics of the robust optimized centrifugal blood pump are investigated. The baseline design is a centrifugal blood pump inspired by a commercial pump used as a left ventricular assist device. The numerical analysis is performed using the SST [Formula: see text] turbulence model and a power-law model of hemolysis. The physical and operational conditions are considered to be uncertain with Beta probability distribution functions. For quantification of the uncertainties, the non-intrusive polynomial chaos method is used, and the assessment of each stochastic parameter's influence on the quantities of interest, that is, the sensitivity analysis, the Sobol' indices are utilized. The primary objective of the present study is to minimize the variation in the pump's efficiency and head coefficient while maintaining the pump's operating point, ensuring that the maximum hemolysis index of the robust optimum design does not exceed that of the baseline design. To this end, robust optimization is carried out via a hybrid evolutionary algorithm. The optimization results clearly show that the robust optimum design is less sensitive to the physical and operational uncertainties. Furthermore, sensitivity analysis is in accordance with theories of turbomachines. Pump efficiency and the maximum hemolysis index are primarily influenced by the variations in blood viscosity, while the head coefficient is mainly affected by rotational speed and flow rate. The pump's velocity field is greatly affected by the mass flow rate in the diffuser regions and the rotational speed in other zones.

如何引用

AMA

Mohamad Sadeq Karimi, Ramin Mohammadi, Mohammad Alihosseini. Robust optimization and stochastic simulation of a centrifugal blood pump under uncertain conditions.. Scientific reports. 2026; doi:10.1038/s41598-026-61510-w.

APA

Mohamad Sadeq Karimi, Ramin Mohammadi, Mohammad Alihosseini (2026). Robust optimization and stochastic simulation of a centrifugal blood pump under uncertain conditions.. Scientific reports. https://doi.org/10.1038/s41598-026-61510-w

GB/T 7714