Spatially Confined Photoinduced Penning Ionization for Highly Sensitive Mass Spectrometry.
Spatially Confined Photoinduced Penning Ionization for Highly Sensitive Mass Spectrometry.
在真空紫外(VUV)灯基光电离质谱(PI-MS)中,轴向安装配置由于光吸收路径较长而具有更高的灵敏度。然而,VUV光照射传输腔内的金属表面会诱导光电子发射;在射频(RF)电场下,光电子电离产生过量背景气体离子,影响PI-MS的定量性能和重复性,尤其是对于光子通量较高的氘灯而言。本研究在电离室内集成了一个尺寸和位置可调的挡板电极,以减少进入传输腔的光泄漏。当直径为3 mm的挡板置于距撇渣器5.4 mm处时,传输腔内产生的背景气体离子强度降至原来的1/506,而三氯乙烯离子强度保持在无挡板时的一半。当进一步向挡板电极施加RF电压时,氯化烃离子强度较SPI模式增强了9倍,此时主导的电离过程为潘宁电离(PenI)。令人惊讶的是,与无挡板时相比,氯化烃离子的强度并未降低,反而最高增加了126%。电场分布和电子动能模拟表明,离子产生区域与高效传输区域的精确重叠对于调控空间受限的PenI以提高灵敏度起着关键作用。该方法实现了低至亚pptv的检测限,即使对于电离能高于光子能量的分子也是如此。
In vacuum ultraviolet (VUV) lamp-based photoionization mass spectrometry (PI-MS), the axially mounted configuration exhibits higher sensitivity due to a longer light absorption path. However, VUV light irradiation of the metal surfaces within the transmission chamber induces photoelectron emission; excessive background gas ions are generated via photoelectron ionization under the radio frequency (RF) electric field, which affects the quantitative performance and repeatability of PI-MS, especially for deuterium lamps with a higher photon flux. In this study, a size- and position-adjustable baffle electrode was integrated into the ionization chamber to reduce the light leakage into the transmission chamber. For a 3 mm-diameter baffle positioned 5.4 mm from the skimmer, the intensity of background gas ions generated in the transmission chamber decreased to 1/506, while the intensity of the trichloroethylene ion was maintained at half that without the baffle. The intensities of chlorinated hydrocarbon ions were enhanced by 9-fold compared with the SPI mode when an RF voltage was further applied to the baffle electrode, under which the dominant ionization process was Penning ionization (PenI). Surprisingly, the intensities of chlorinated hydrocarbon ions did not decrease but increased by up to 126% compared to those without the baffle. Simulation of the electric field distribution and electron kinetic energy highlights that the precise overlap of the ion generation region and efficient transmission region plays a pivotal role in the regulation of spatially confined PenI toward sensitivity improvement. This approach has achieved limits of detection down to sub-pptv, even for molecules with ionization energies higher than the photon energy.
Yi Yu, Hang Li, Zhenyuan Zhang, Zhihao Zhang, Tianhao Huang, Wei Huang, et al. Spatially Confined Photoinduced Penning Ionization for Highly Sensitive Mass Spectrometry.. Analytical chemistry. 2026; doi:10.1021/acs.analchem.6c01680.
Yi Yu, Hang Li, Zhenyuan Zhang, Zhihao Zhang, Tianhao Huang, Wei Huang, et al (2026). Spatially Confined Photoinduced Penning Ionization for Highly Sensitive Mass Spectrometry.. Analytical chemistry. https://doi.org/10.1021/acs.analchem.6c01680
Yi Yu, Hang Li, Zhenyuan Zhang, Zhihao Zhang, Tianhao Huang, Wei Huang, et al. Spatially Confined Photoinduced Penning Ionization for Highly Sensitive Mass Spectrometry.[J]. Analytical chemistry, 2026 doi:10.1021/acs.analchem.6c01680.
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出