一种计算3≤Zeff≤20的生物医学对象物理密度的算法:使用临床光子计数计算机断层扫描设备的应用研究

{"_":"An algorithm to calculate physical density of biomedical objects having 3 ≤ Z ≤ 20: an application study using clinical photon-counting computed tomography equipment.","sub":["eff"]}

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

近年来,光子计数计算机断层扫描(PC-CT)设备已可用于临床X射线诊断,开发用于生成各种定量图像的分析算法也备受关注。本研究提出一种算法,用于计算有效原子序数(Z)在3-20范围内的生物对象的有效物理密度(ρ)。在我们的流程中,当将相互作用截面(σs)拟合到由虚拟单色图像(VMI)测定的实测线性衰减系数(μs)时,即可确定ρ。在此过程中,Z被预先分析,并将该信息反馈至ρ分析。我们流程的一个关键特征是不需要基于特定物质的校准过程,这使我们能够在不做任何假设的情况下分析任何物质。为验证所提出算法的有效性,我们使用临床PC-CT扫描仪进行了实验。扫描了多能CT体模、水体模、自制低密度体模、食品样品和胸部体模。结果表明,我们的流程可以计算水的ρ,不确定度为4.5%。本研究结果的学术意义在于证明,先前从CT值获得的对比度贡献因素可以利用Z和ρ的信息进行物理解释。这种分析无法用传统单能CT扫描仪完成,而用PC-CT和双能CT扫描仪则可以实现。我们的新流程有望在X射线诊断中提供有用的附加信息。

Abstract

Recently, photon-counting computed tomography (PC-CT) devices have become clinically available for X-ray diagnosis, and developments of analysis algorithms to generate various quantitative images have focused attention. This study proposes an algorithm to calculate the effective physical density (ρ) of biological objects having an effective atomic number (Z) of 3-20. In our procedure, ρ was determined when fitting the interaction cross sections (σs) to the measured linear attenuation coefficients (μs), which were determined from virtual monochromatic images (VMIs). In this process, the Z was analyzed in advance, and this information was fed back to the ρ analysis. A key feature of our procedure is that it does not require calibration processes based on specific substances, which allow us to analyze any substance without making assumptions. To validate the availability of the proposed algorithm, we conducted an experiment using a clinical PC-CT scanner. A multi-energy CT phantom, a water phantom, an in-house low-density phantom, food samples, and a chest phantom were scanned. The results showed that our procedure can calculate ρ of water with an uncertainty of 4.5%. The academic significance of the present research results lies in demonstrating that the factors contributing to contrast, previously obtained from CT values, can be physically interpreted using information from Z and ρ. This analysis cannot be performed with conventional single-energy CT scanners and is possible with PC-CT and dual-energy CT scanners. Our novel procedure is expected to provide useful additional information in X-ray diagnosis.

如何引用

AMA

Hiroaki Hayashi, Rina Nishigami, Takashi Asahara, Natsumi Kimoto, Daiki Kobayashi, Mana Mitani, et al. {"_":"An algorithm to calculate physical density of biomedical objects having 3 ≤ Z ≤ 20: an application study using clinical photon-counting computed tomography equipment.","sub":["eff"]}. Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine. 2026; doi:10.1016/j.apradiso.2026.112768.

APA

Hiroaki Hayashi, Rina Nishigami, Takashi Asahara, Natsumi Kimoto, Daiki Kobayashi, Mana Mitani, et al (2026). {"_":"An algorithm to calculate physical density of biomedical objects having 3 ≤ Z ≤ 20: an application study using clinical photon-counting computed tomography equipment.","sub":["eff"]}. Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine. https://doi.org/10.1016/j.apradiso.2026.112768

GB/T 7714

Hiroaki Hayashi, Rina Nishigami, Takashi Asahara, Natsumi Kimoto, Daiki Kobayashi, Mana Mitani, et al. {"_":"An algorithm to calculate physical density of biomedical objects having 3 ≤ Z ≤ 20: an application study using clinical photon-counting computed tomography equipment.","sub":["eff"]}[J]. Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine, 2026 doi:10.1016/j.apradiso.2026.112768.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI

Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine 的其他文献