皮肌炎对无创、定量的肌无力评估提出了临床需求,而通过分解表面肌电信号(sEMG)中的神经运动信息可实现这一目标。然而,缺乏与病理性干燥皮肤相容的稳定电极界面阻碍了该方法的实现。在此,我们开发了一种顺应性神经运动电极界面,通过分子工程克服了这一障碍,其具有超软模量(约2.13 kPa)、低阻抗(1 kHz时约0.9 kΩ)以及7天内稳定的电学保持性能。该界面能够从皮肌炎患者中采集高保真sEMG信号,基线噪声低(均方根值< 2 µV)且通道间信号相关性高,从而实现对神经运动生物标志物的稳健分解,包括运动单位数量、动作电位幅度和发放频率。在临床上,我们的多生物标志物定量策略与金标准评估显示出高度一致性。此外,我们的方法能够检测患者间的细微差异,如早期募集模式,从而在临床金标准之外实现个性化治疗。这项工作为精准肌无力评估建立了一个变革性平台,克服了针极肌电图和主观临床量表的局限性。

Abstract

Dermatomyositis creates a clinical demand for non-invasive, quantitative muscle weakness assessment, which can be achieved by decomposing neuromotor information from surface electromyography (sEMG). However, the absence of a stable electrode interface compatible with pathologically xerotic skin has hindered this approach. Here, we develop a conformable neuromotor electrode interface that overcomes this via molecular engineering, exhibiting an ultrasoft modulus (∼ 2.13 kPa), low impedance (∼ 0.9 kΩ at 1 kHz), and stable electrical retention over 7 days. This interface enables the acquisition of high-fidelity sEMG from dermatomyositis patients with low baseline noise (< 2 µV root mean square) and high inter-channel signal correlation, allowing robust decomposition of neuromotor biomarkers including motor unit count, action potential amplitude, and firing rate. Clinically, our multi-biomarker quantification strategy shows strong agreement with gold standard assessments. Moreover, our approach can detect subtle interpatient variations like early recruitment patterns, enabling personalized treatment beyond the clinical gold standard. This work establishes a transformative platform for precise muscle weakness evaluation, overcoming the limitations of needle electromyography and subjective clinical scales.

如何引用

AMA

Yue Zhang, Jie Cao, Xinzhi Xu, Zhiwei Chen, Yaqing Li, Junchang Wang, et al. A Xerotic Skin-Compatible Neuromotor Electrode Interface for Non-Invasive Muscle Weakness Evaluation in Dermatomyositis.. Advanced materials (Deerfield Beach, Fla.). 2026; doi:10.1002/adma.73874.

APA

Yue Zhang, Jie Cao, Xinzhi Xu, Zhiwei Chen, Yaqing Li, Junchang Wang, et al (2026). A Xerotic Skin-Compatible Neuromotor Electrode Interface for Non-Invasive Muscle Weakness Evaluation in Dermatomyositis.. Advanced materials (Deerfield Beach, Fla.). https://doi.org/10.1002/adma.73874

GB/T 7714

Yue Zhang, Jie Cao, Xinzhi Xu, Zhiwei Chen, Yaqing Li, Junchang Wang, et al. A Xerotic Skin-Compatible Neuromotor Electrode Interface for Non-Invasive Muscle Weakness Evaluation in Dermatomyositis.[J]. Advanced materials (Deerfield Beach, Fla.), 2026 doi:10.1002/adma.73874.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI

Advanced materials (Deerfield Beach, Fla.) 的其他文献