一种基于反射率的多模式可穿戴光电容积描记(PPG)传感器

A Reflectance-based multimodal wearable photoplethysmography (PPG) sensor.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

本文介绍了一种基于反射式的可穿戴光电容积脉搏波(PPG)传感器ChromaSense,旨在实现对不同肤色人群生命体征的稳健监测,包括血氧饱和度(SpO2)、脉率和呼吸率。传统基于PPG的传感器易受黑色素引起的光学衰减影响,这可能导致测量误差,尤其是在肤色较深的人群中。为解决这一局限,ChromaSense集成了光谱比色传感技术,用于估算皮肤反射率并自适应调节LED驱动电流。该系统在涵盖Fitzpatrick肤色量表的50名参与者队列中进行了评估。对于SpO2,该设备相对于美国食品药品监督管理局批准的参考脉搏血氧仪(Masimo MightySAT)达到了1.36%的均方根准确度,满足ISO 80601-2-61:2013标准中≤3.5%的要求。Bland-Altman分析显示平均偏差为-0.27%,95%一致性界限为-2.91%至2.36%。对于脉率估算,84%的测量值在参考值的±10%以内;对于呼吸率估算,90%的测量值在参考值的±5次/分钟以内。单因素方差分析(ANOVA)显示,不同Fitzpatrick肤色类型之间的SpO2测量误差无统计学显著差异(p = 0.518)。这些结果表明,所提出的多模态传感策略在受评估队列中减轻了肤色依赖性变异,并能够在不同用户之间实现一致的生理测量性能。

Abstract

This paper presents ChromaSense, a reflectance-based wearable photoplethysmography (PPG) sensor designed for robust monitoring of vital signs, including blood oxygen saturation (SpO2), pulse rate, and respiration rate across diverse skin tones. Conventional PPG-based sensors are susceptible to melanin-induced optical attenuation, which can contribute to measurement error, particularly in individuals with darker skin pigmentation. To address this limitation, ChromaSense integrates spectral colorimetric sensing to estimate skin reflectance and adaptively adjust LED drive current. The system was evaluated on a cohort of 50 participants spanning the Fitzpatrick skin tone scale. For SpO2, the device achieved a root-mean-square accuracy of 1.36% relative to a Food and Drug Administration-cleared reference pulse oximeter (Masimo MightySAT), satisfying the ISO 80601-2-61:2013 standard requirement of ≤3.5%. Bland-Altman analysis showed a mean bias of -0.27% with 95% limits of agreement from -2.91% to 2.36%. For pulse rate estimation, 84% of measurements were within ±10% of the reference values, and for respiration rate estimation, 90% of measurements were within ±5 breaths per minute of the reference values. A one-way analysis of variance (ANOVA) showed no statistically significant differences in SpO2 measurement error across Fitzpatrick skin types (p = 0.518). These results suggest that the proposed multimodal sensing strategy mitigates skin-tone-dependent variability within the evaluated cohort and enables consistent physiological measurement performance across diverse users.

如何引用

AMA

Ravi Durbha, Valencia Koomson. A Reflectance-based multimodal wearable photoplethysmography (PPG) sensor.. The Review of scientific instruments. 2026; doi:10.1063/5.0304023.

APA

Ravi Durbha, Valencia Koomson (2026). A Reflectance-based multimodal wearable photoplethysmography (PPG) sensor.. The Review of scientific instruments. https://doi.org/10.1063/5.0304023

GB/T 7714

Ravi Durbha, Valencia Koomson. A Reflectance-based multimodal wearable photoplethysmography (PPG) sensor.[J]. The Review of scientific instruments, 2026 doi:10.1063/5.0304023.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI

The Review of scientific instruments 的其他文献