一种用于非侵入性单次试验检测皮质群体尖峰的超灵敏多通道MEG系统
An ultra-sensitive multi-channel MEG system for the non-invasive single-trial detection of cortical population spikes.
An ultra-sensitive multi-channel MEG system for the non-invasive single-trial detection of cortical population spikes.
我们描述了一套高灵敏度12通道脑磁图(MEG)系统的设计与性能,该系统将在磁屏蔽室内运行。该记录系统的目标是在单次试验水平上测量由正中神经电刺激诱发的高频(450-850 Hz)躯体感觉诱发反应(hfSERs)的空间分辨信号。该系统采用在超低噪声杜瓦瓶中运行的电流传感超导量子干涉器件(SQUIDs)。传感器头由9个多匝六边形信号磁强计、2个单匝圆形参考磁强计和1个圆形轴向梯度计组成。鉴于杜瓦瓶尺寸的限制,在最小化磁场噪声与各MEG通道的空间分布之间找到了平衡。使用目前由德国联邦物理技术研究院制造的SQUIDs,预期最低白噪声性能为[公式:见原文],在感兴趣频段内实现的噪声性能为[公式:见原文]。此外,对hfSERs神经元磁场的模拟预测,我们的系统峰峰强度为35 fT,与此前的记录结果一致。我们还提出了提高MEG设备动态范围的措施,并展示了在体测量结果,证明在单次试验水平上可对振幅为[公式:见原文]的hfSERs进行空间分辨检测。
We describe the design and performance of a highly sensitive 12-channel magnetoencephalography (MEG) system to be operated in a magnetically shielded room. The aim of the recording system is to measure spatially resolved high frequency (450-850 Hz) somatosensory evoked responses (hfSERs), as elicited by electrostimulation of the median nerve, at a single-trial level. The system uses current sensing superconducting quantum interference devices (SQUIDs) operated in an ultra-low noise dewar. The sensor head consists of 9 multi-turn, hexagonal signal magnetometers, 2 single-turn, circular reference magnetometers and one circular axial gradiometer. Given limitations in the dewar's size, a balance was found between minimising the magnetic field noise and the spatial distribution of the individual MEG channels. Using SQUIDs currently made at the Physikalisch-Technische Bundesanstalt, the minimum white noise performance is expected to be [Formula: see text], with the achieved noise performance in the frequency band of interest of [Formula: see text]. Additionally, simulations of the neuronal magnetic field of the hfSERs predicted a peak-to-peak strength of 35 fT for our system, in accordance with previous recordings. We also present measures to increase the dynamic range of our MEG device and show in vivo measurement results which demonstrate spatially resolved detection of the hfSERs with amplitudes of [Formula: see text] at the single-trial level.
Jim Barnes, Soudabeh Arsalani, Gunnar Waterstraat, Gabriel Curio, Lukasz Radzinski, Jens Haueisen, et al. An ultra-sensitive multi-channel MEG system for the non-invasive single-trial detection of cortical population spikes.. Scientific reports. 2026; doi:10.1038/s41598-026-50113-0.
Jim Barnes, Soudabeh Arsalani, Gunnar Waterstraat, Gabriel Curio, Lukasz Radzinski, Jens Haueisen, et al (2026). An ultra-sensitive multi-channel MEG system for the non-invasive single-trial detection of cortical population spikes.. Scientific reports. https://doi.org/10.1038/s41598-026-50113-0
Jim Barnes, Soudabeh Arsalani, Gunnar Waterstraat, Gabriel Curio, Lukasz Radzinski, Jens Haueisen, et al. An ultra-sensitive multi-channel MEG system for the non-invasive single-trial detection of cortical population spikes.[J]. Scientific reports, 2026 doi:10.1038/s41598-026-50113-0.
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出