用于乳腺癌症成像的光计数CT中的光谱性能的体模评估
Phantom evaluation of spectral performance in photon-counting CT for breast cancer imaging.
Phantom evaluation of spectral performance in photon-counting CT for breast cancer imaging.
对比增强是检测乳腺恶性肿瘤最敏感的指标。由于软组织对比度较低,计算机断层扫描(CT)在乳腺肿瘤局部区域分期中的作用有限。本研究旨在利用对比增强乳腺摄影(CEM)体模评估并优化临床光子计数计算机断层扫描(PCCT)在乳腺癌成像中的性能,并将其成像性能与双源双能量CT(DS-DECT)进行比较。将含有模拟乳腺病灶的CEM体模置于拟人胸部体模上,使用临床PCCT系统以120 kV多能量模式进行扫描,并使用DS-DECT系统以70/Sn150 kV和90/Sn150 kV两组管电压对进行扫描。PCCT扫描仪变量包括扫描模式[标准分辨率(SR)和超高分辨率(UHR)]、视野(FOV)大小(大和小)、矩阵大小(512和1024),以及用于分析的图像类型[低能量阈值图像、50、60和70 keV下的虚拟单能图像(VMI)以及碘图]。使用置于体模内含碘病灶及背景上的圆形感兴趣区(ROI)进行定量分析。对于每个ROI,在中央五层图像和三次独立扫描中测量平均CT值或碘浓度及标准差。在所有PCCT配置和图像类型中评估对比噪声比(CNR)和圆形度,并与DS-DECT获得的结果进行比较。在所有PCCT配置中,采用小FOV的UHR模式配合512或1024矩阵在50 keV VMI下,在所有碘浓度中实现了最高的综合CNR。此外,采用512矩阵的UHR模式配合小或大FOV产生了最高的综合圆形度值。与DECT扫描衍生的50 keV VMI相比,最佳PCCT配置实现了更高的CNR,以及更高或相当的圆形度。本体模研究表明,使用UHR模式、低keV VMI、常规矩阵大小和专用重建FOV可实现PCCT用于潜在乳腺癌成像的最佳能谱性能,优于DECT。
Contrast enhancement is the most sensitive indicator for detecting breast malignancies. Computed tomography (CT) has had a limited role for the locoregional staging of breast tumors due to low soft tissue contrast. To evaluate and optimize the performance of clinical photon-counting computed tomography (PCCT) for breast cancer imaging using a contrast-enhanced mammography (CEM) phantom and to compare its imaging performance with dual-source dual-energy CT (DS-DECT). A CEM phantom containing simulated breast lesions was positioned on an anthropomorphic thoracic phantom and scanned using a clinical PCCT system at 120 kV in multi-energy mode and a DS-DECT system with two kV pairs of 70/Sn150 kV and 90/Sn150 kV. PCCT scanner variables included scan mode [standard resolution (SR) and ultra-high resolution (UHR)], field of view (FOV) size (large and small), matrix size (512 and 1024), and type of image used for analysis [low-energy threshold images, virtual monoenergetic images (VMIs) at 50, 60, and 70 keV, and iodine maps]. Quantitative analysis was performed using circular regions of interest (ROIs) placed on iodine-containing lesions and background within the phantom. For each ROI, mean CT numbers or iodine concentrations and standard deviations were measured across the central five slices and three independent scans. Contrast-to-noise ratio (CNR) and circularity were evaluated across all PCCT configurations and image types and compared with those obtained from DS-DECT. Among all PCCT configurations, the UHR mode with a small FOV and either a 512 or 1024 matrix at 50 keV VMI achieved the highest combined CNR across all iodine concentrations. Additionally, the UHR mode with a 512 matrix and either small or large FOV yielded the highest combined circularity values. The optimal PCCT configuration achieved higher CNR, and higher or comparable circularity compared with 50 keV VMIs derived from DECT scans. This phantom study demonstrated that optimal spectral performance for potential breast cancer imaging with PCCT is achieved using UHR mode, low-keV VMIs, a regular matrix size, and dedicated reconstruction FOVs, outperforming DECT.
Liqiang Ren, Yin Xi, Lakshmi Ananthakrishnan, Todd Soesbe, Matthew Lewis, Stephen Seiler, et al. Phantom evaluation of spectral performance in photon-counting CT for breast cancer imaging.. Medical physics. 2026; doi:10.1002/mp.70472.
Liqiang Ren, Yin Xi, Lakshmi Ananthakrishnan, Todd Soesbe, Matthew Lewis, Stephen Seiler, et al (2026). Phantom evaluation of spectral performance in photon-counting CT for breast cancer imaging.. Medical physics. https://doi.org/10.1002/mp.70472
Liqiang Ren, Yin Xi, Lakshmi Ananthakrishnan, Todd Soesbe, Matthew Lewis, Stephen Seiler, et al. Phantom evaluation of spectral performance in photon-counting CT for breast cancer imaging.[J]. Medical physics, 2026 doi:10.1002/mp.70472.