对一种广泛使用的磁带记录仪(Oxford Medilog)和心律失常计算机(Reynolds Pathfinder)的频率响应及走纸速度稳定性进行了评估,并将动态心电图(ambulatory ECG)上的时间和振幅测量值与常规心电图(conventional ECG)中的相应测量值进行了比较。总带宽为0.15~33 Hz(下降3 dB),平均速度误差为+2%±1.8%。两种方法测得的P-Q间期和Q-T间期差异无统计学意义,但动态心电图中的波幅显著降低。动态心电图中QRS波群的界限常模糊不清,临界性波群仅凭QRS宽度难以归类为正常或异常。人为性J点压低是动态心电图中的常见表现,而水平型或下斜型ST段压低则能被忠实再现。
The frequency response and speed stability of a widely used tape recorder (Oxford Medilog) and arrhythmia computer (Reynolds Pathfinder) was evaluated and measurement in time and amplitude on the ambulatory ECG compared with correspondent measurement in the conventional ECG. The overall bandwidth was 0.15--33Hz (3 dB down) and the mean speed error +2% +/- 1.8%. There was no statistically significant difference between P-Q and Q-T intervals measured by the 2 methods, but a significant reduction in wave amplitudes in the ambulatory ECG. The demarcation of QRS complexes was often blurred in the ambulatory ECG and borderline complexes difficult to classify as normal or abnormal on QRS width alone. Artifical J point depression was a common finding in the ambulatory ECT, whereas horizontal or downsloping ST segment depression was faithfully reproduced.
P Bjerregaard. Distortions in the ECG caused by instruments for ambulatory electrocardiography.. Biotelemetry and patient monitoring. 1980;
P Bjerregaard (1980). Distortions in the ECG caused by instruments for ambulatory electrocardiography.. Biotelemetry and patient monitoring.
P Bjerregaard. Distortions in the ECG caused by instruments for ambulatory electrocardiography.[J]. Biotelemetry and patient monitoring, 1980
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出