运动试验中的12导联心电图。一个误导性的基线?
The 12-lead electrocardiogram in exercise testing. A misleading baseline?
The 12-lead electrocardiogram in exercise testing. A misleading baseline?
Mason和Likar报道了一种改良的心电图导联系统,据称该系统在不改变所记录导联构型的情况下,基本消除了运动过程中的伪差。该系统可在运动期间及运动后连续监测全部12个导联,并已成为其他系统比较的标准。然而,我们发现,使用Mason-Likar改良法获得的基线运动心电图与同步记录的标准心电图之间可能存在重要差异。虽然这些差异显然不影响试验结果的判读,但它们可能在基线运动心电图上前置性梗死的证据产生或掩盖,因此不能将其视为可与标准心电图互换。
Mason and Likar reported a modified ECG lead system that largely eliminates intraexercise artifact without reportedly altering the configuration of the recorded leads. This system allows continuous monitoring of all 12 leads during and after exercise and has become the standard by which other systems are compared. We have found, however, that important changes may be seen between the baseline exercise ECG using the Mason-Likar modification and a simultaneous standard ECG. While these differences do not apparently affect the interpretation of the test, they may produce or obscure evidence of antecedent infarction on the baseline exercise ECG and thus preclude it from being considered interchangeable with a standard ECG.
J P Kleiner, W P Nelson, M J Boland. The 12-lead electrocardiogram in exercise testing. A misleading baseline?. Archives of internal medicine. 1978;
J P Kleiner, W P Nelson, M J Boland (1978). The 12-lead electrocardiogram in exercise testing. A misleading baseline?. Archives of internal medicine.
J P Kleiner, W P Nelson, M J Boland. The 12-lead electrocardiogram in exercise testing. A misleading baseline?[J]. Archives of internal medicine, 1978