目视评估与溶血指数在常见生化检测中检测溶血的比较:对临床实验室实践中小样本拒收的意义

Comparison of Visual Assessment and Hemolysis Index in Detecting Hemolysis across Common Biochemical Assay: Implications for Sample Rejection in Clinical Laboratory Practice.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

溶血是临床实验室中常见的分析前干扰因素,可影响高达4%的血液样本。它可显著影响检测结果的准确性,可能导致误诊和不恰当的治疗。通过比较视觉评估与机器溶血指数(HIs)来量化溶血对常规临床生化参数的影响。本横断面研究于2021年至2023年在AIIMS,Kalyani进行。使用残余血清样本评估机械诱导的溶血对Vitros 4600分析仪测定的40项生化参数的影响。开发了半定量视觉分级量表,并与分析仪的定量HI阈值进行比较。利用医院电子健康记录数据库分析了视觉方法与自动化方法之间样本拒收率的差异。数据分析包括计算拒收率并比较每种分析物视觉方法与基于HI方法之间的差异。视觉评估将大多数样本归类为“无溶血”(91.86%-100%),轻度溶血(1.43%-9.09%)常见,重度溶血(0.41%-0.62%)罕见。基于HI的检测显示,钾(高1.58%)、磷酸盐(1.07%)和肝酶等分析物的拒收率较高。相反,肌酐(+0.75%)、尿素(+0.79%)和电解质的拒收率较低。频繁检测的分析物中注意到超过0.5%的差异,引发了对特定参数视觉评估可靠性的担忧。通过使HI阈值与分析灵敏度和临床需求相一致来优化溶血评估方案,有望提高实验室结果准确性和患者护理质量。

Abstract

Hemolysis is a common preanalytical interference in clinical laboratories, affecting up to 4% of blood samples. It can significantly impact the accuracy of test results, potentially leading to misdiagnosis and inappropriate treatment. To quantify the effects of hemolysis on routine clinical biochemistry parameters by comparing visual assessment with machine-based hemolysis indices (HIs). This cross-sectional study was conducted at AIIMS, Kalyani, from 2021 to 2023. Residual serum samples were used to assess the impact of mechanically induced hemolysis on 40 biochemical parameters measured using a Vitros 4600 analyzer. A semiquantitative visual grading scale was developed and compared to the analyzer's quantitative HI thresholds. Discrepancies in sample rejection rates between visual and automated methods were analyzed using the hospital's electronic health record database. Data analysis included calculating rejection rates and comparing differences between visual and HI-based methods for each analyte. Visual assessment categorized most samples as "no hemolysis" (91.86%-100%), with mild hemolysis (1.43%-9.09%) being common and severe hemolysis (0.41%-0.62%) rare. HI-based detection showed higher rejection rates for analytes such as potassium (1.58% higher), phosphate (1.07%), and liver enzymes. Conversely, lower rejection rates were observed for creatinine (+0.75%), urea (+0.79%), and electrolytes. Discrepancies exceeding 0.5% were noted for frequently tested analytes, raising the concerns about the reliability of visual assessment for specific parameters. Optimizing hemolysis assessment protocols by aligning HI thresholds with analytical sensitivity and clinical needs shows promise for improving laboratory result accuracy and patient care.

如何引用

AMA

Sibasish Sahoo, Kaushik Mukhopadhyay, Atanu Kumar Dutta, Amit Pal, Kalyan Goswami. Comparison of Visual Assessment and Hemolysis Index in Detecting Hemolysis across Common Biochemical Assay: Implications for Sample Rejection in Clinical Laboratory Practice.. Annals of African medicine. 2026; doi:10.4103/aam.aam_211_25.

APA

Sibasish Sahoo, Kaushik Mukhopadhyay, Atanu Kumar Dutta, Amit Pal, Kalyan Goswami (2026). Comparison of Visual Assessment and Hemolysis Index in Detecting Hemolysis across Common Biochemical Assay: Implications for Sample Rejection in Clinical Laboratory Practice.. Annals of African medicine. https://doi.org/10.4103/aam.aam_211_25

GB/T 7714

Sibasish Sahoo, Kaushik Mukhopadhyay, Atanu Kumar Dutta, Amit Pal, Kalyan Goswami. Comparison of Visual Assessment and Hemolysis Index in Detecting Hemolysis across Common Biochemical Assay: Implications for Sample Rejection in Clinical Laboratory Practice.[J]. Annals of African medicine, 2026 doi:10.4103/aam.aam_211_25.