阿尔茨海默病血浆生物标志物化学发光免疫测定小组的分析验证和临床性能:一项初步研究
Analytical validation and clinical performance of a chemiluminescence immunoassay panel for plasma biomarkers in Alzheimer's disease: a preliminary study.
Analytical validation and clinical performance of a chemiluminescence immunoassay panel for plasma biomarkers in Alzheimer's disease: a preliminary study.
阿尔茨海默病(AD)构成日益严峻的全球健康挑战,凸显了对可靠且微创诊断工具的迫切需求。本研究旨在开发并全面验证一种自动化化学发光免疫分析法(CLIA),用于同时定量检测六种核心血浆AD生物标志物:Aβ1-40、Aβ1-42、p-Tau181、p-Tau217、GFAP和NfL。采用全自动化CLIA平台(Shine i2910,Vazyme,中国)进行生物标志物检测。共纳入来自北京天坛医院确诊住院患者的87份血浆样本,分为三个队列:AD、其他痴呆(OD)和轻度认知障碍(MCI)。该检测方法的分析性能,包括精密度、准确度、线性和检测限,依据临床和实验室标准协会(CLSI)指南(EP15-A3、EP09-A3、EP06-A2、EP17-A2)进行验证。该CLIA检测组合实现了稳健的分析指标,所有六种生物标志物的批内变异系数(CV)< 3%、批间CV < 5%、相对偏差< ±10%、线性相关系数(R)> 0.99,并具有亚皮克级检测灵敏度。临床样本分析显示,与OD和MCI患者相比,AD患者血浆p-Tau181和p-Tau217浓度显著升高,同时Aβ1-42/Aβ1-40显著降低(p < 0.05)。与MCI患者相比,AD和OD患者的GFAP和NfL浓度均显著升高(p < 0.05),而AD与OD患者之间无显著差异。受试者工作特征(ROC)曲线分析显示,所有生物标志物的曲线下面积(AUC)> 0.7,支持AD的辅助鉴别诊断,并证明了该检测方法的临床区分能力。所开发的全自动化CLIA平台能够以令人满意的分析性能快速、高性能地定量检测六种核心血浆AD生物标志物。作为一项无匹配健康对照的单中心探索性研究,本研究仅提供了分析验证和初步的临床区分证据。它为临床研究条件下的辅助AD筛查提供了一种经济有效的微创替代方案,但仍需大型多中心队列来验证其真实世界大规模筛查价值。
Alzheimer's disease (AD) poses a growing global health challenge, highlighting the urgent need for reliable and minimally invasive diagnostic tools. This study aimed to develop and comprehensively validate an automated chemiluminescence immunoassay (CLIA) for the simultaneous quantification of six core plasma AD biomarkers: Aβ1-40, Aβ1-42, p-Tau181, p-Tau217, GFAP and NfL. A fully automated CLIA platform (Shine i2910, Vazyme, China) was utilized for biomarker detection. A total of 87 plasma samples from hospitalized patients with confirmed diagnosis at Beijing Tiantan Hospital were stratified into three cohorts: AD, Other Dementia (OD), and Mild Cognitive Impairment (MCI). The analytical performance of the assay, including precision, accuracy, linearity, and detection limits, was validated in accordance with Clinical and Laboratory Standards Institute (CLSI) guidelines (EP15-A3, EP09-A3, EP06-A2, EP17-A2). This CLIA panel achieved robust analytical metrics, with within-run coefficient of variation (CV) < 3%, between-run CV < 5%, relative bias < ±10%, and linear correlation coefficient (R) > 0.99 for all six biomarkers, along with sub-picogram level detection sensitivity. Clinical sample analysis revealed that AD patients had significantly higher plasma concentrations of p-Tau181 and p-Tau217, along with a significantly reduced Aβ1-42/Aβ1-40, compared with OD and MCI patients (p < 0.05). GFAP and NfL concentrations were significantly elevated in both AD and OD patients relative to the MCI patients (p < 0.05), with no significant difference between AD and OD patients. Receiver operating characteristic (ROC) curve analysis demonstrated an area under the curve (AUC) > 0.7 for all biomarkers, supporting auxiliary differential diagnosis of AD and demonstrating the assay's clinical discriminatory ability. The developed fully automated CLIA platform enables rapid, high-performance quantification of six core plasma AD biomarkers with satisfactory analytical performance. As a single-center exploratory study without matched healthy controls, this study only provides analytical validation and preliminary clinical discriminatory evidence. It offers a cost-effective, minimally invasive alternative for auxiliary AD screening under clinical research conditions, while large multi-center cohorts are required to verify its real-world large-scale screening value.
Ximeng Chen, Wencan Jiang, Lijuan Wang, Bingqing Han, Danwei Yu, Guoge Li, et al. Analytical validation and clinical performance of a chemiluminescence immunoassay panel for plasma biomarkers in Alzheimer's disease: a preliminary study.. Clinica chimica acta; international journal of clinical chemistry. 2026; doi:10.1016/j.cca.2026.121300.
Ximeng Chen, Wencan Jiang, Lijuan Wang, Bingqing Han, Danwei Yu, Guoge Li, et al (2026). Analytical validation and clinical performance of a chemiluminescence immunoassay panel for plasma biomarkers in Alzheimer's disease: a preliminary study.. Clinica chimica acta; international journal of clinical chemistry. https://doi.org/10.1016/j.cca.2026.121300
Ximeng Chen, Wencan Jiang, Lijuan Wang, Bingqing Han, Danwei Yu, Guoge Li, et al. Analytical validation and clinical performance of a chemiluminescence immunoassay panel for plasma biomarkers in Alzheimer's disease: a preliminary study.[J]. Clinica chimica acta; international journal of clinical chemistry, 2026 doi:10.1016/j.cca.2026.121300.
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出