通过混合离子推断实现离子选择性电极的自动化工作流程

Automated Workflow for Ion-Selective Electrodes via Mixed-Ion Inference.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

从盐湖卤水中选择性电化学提锂需要反复进行溶液置换,并在捕获和释放步骤中定量追踪Li/Na组成。在此,我们开发了一个自动化平台,将基于锰酸锂的电化学提锂、可编程流体处理和成对的Li/Na离子选择性电极分析相结合。为实现在混合溶液中的定量,我们将已有的纯离子、分离溶液和固定干扰表征方法与一个在混合LiCl/NaCl组成范围内训练的、具有物理可解释性的双传感器模型相结合。该模型不是通过单一选择性系数来评估传感器性能,而是通过在实验校准的组成范围内同时回收Li和Na浓度来直接检验。该平台在重复提取循环中自动完成溶液输送、电化学捕获、洗涤、释放、样品收集以及随后的在线电位定量,从而实现循环分辨的Li/Na分离性能测定。这一应用驱动的工作流程将自动化电化学分离与混合离子组成推断相连接,并为锂回收过程的自主评估与优化提供了一条透明的路径。

Abstract

Selective electrochemical extraction of Li from brines requires repeated solution exchange and quantitative tracking of Li/Na composition across capture and release steps. Here, we develop an automated platform that couples lithium manganese oxide-based electrochemical Li extraction, programmable fluid handling, and paired Li/Na ion-selective electrode analysis. To enable quantification in mixed solutions, established pure-ion, separate-solution, and fixed-interference characterization is combined with a physically interpretable dual-sensor model trained across mixed LiCl/NaCl compositions. Rather than evaluating sensor performance through a single selectivity coefficient, the model is tested directly through simultaneous recovery of Li and Na concentrations within the experimentally calibrated composition domain. The platform automates solution delivery, electrochemical capture, washing, release, sample collection, and subsequent at-line potentiometric quantification over repeated extraction cycles, enabling cycle-resolved determination of Li/Na separation performance. This application-driven workflow connects automated electrochemical separation with mixed-ion composition inference and provides a transparent route toward autonomous evaluation and optimization of Li recovery processes.

如何引用

AMA

Mamta Dagar, Thomas D Pope, Ulan Sarolia, Hang Ren. Automated Workflow for Ion-Selective Electrodes via Mixed-Ion Inference.. Sensors and actuators. B, Chemical. 2026; doi:10.1016/j.snb.2026.140650.

APA

Mamta Dagar, Thomas D Pope, Ulan Sarolia, Hang Ren (2026). Automated Workflow for Ion-Selective Electrodes via Mixed-Ion Inference.. Sensors and actuators. B, Chemical. https://doi.org/10.1016/j.snb.2026.140650

GB/T 7714

Mamta Dagar, Thomas D Pope, Ulan Sarolia, Hang Ren. Automated Workflow for Ion-Selective Electrodes via Mixed-Ion Inference.[J]. Sensors and actuators. B, Chemical, 2026 doi:10.1016/j.snb.2026.140650.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI