Effect of a proprietary motion artifact correction algorithm on CBCT spatial resolution: an in vitro phantom study.
Effect of a proprietary motion artifact correction algorithm on CBCT spatial resolution: an in vitro phantom study.
评估专有运动伪影校正算法在体外模体模型中对CBCT空间分辨率的影响。使用ProMax 3D设备,将点扩散函数模体安装于电动转台上,共采集50次CBCT扫描。测试了10种方案,包括无运动及诱导运动至9度,分别在有和无校正算法激活的条件下进行。采用半高全宽(FWHM)、表面图、快速傅里叶变换、配对比较和Bland-Altman分析评估空间分辨率。运动增加了FWHM值并使点扩散函数发生畸变,随着运动幅度增大,不规则程度加重。校正算法在部分运动方案中产生部分改善,尤其是在较低运动幅度下,但其效果在所有运动条件和测量方向上并不一致。在无运动方案中,激活算法引入了轻微的不锐利度并增加了高频域的异质性。患者运动降低CBCT空间分辨率。该专有校正算法在特定条件下可提供部分改善,尤其是在较低运动幅度下,但不能替代图像采集过程中对患者进行仔细固定的必要性,在不太可能发生运动时应谨慎使用。
To evaluate the effect of a proprietary motion artifact correction algorithm on CBCT spatial resolution using an in vitro phantom model. Fifty CBCT scans were acquired with a ProMax 3D unit using a point spread function phantom mounted on a motorized turntable. Ten protocols were tested, including no motion and induced motion up to 9 degrees, with and without activation of the correction algorithm. Spatial resolution was assessed using full width at half maximum (FWHM), surface plots, fast Fourier transforms, pairwise comparisons, and Bland-Altman analysis. Motion increased FWHM values and distorted the point spread function, with greater irregularity as motion magnitude increased. The correction algorithm produced partial improvement in selected motion protocols, particularly at lower motion magnitudes, but the effect was inconsistent across all motion conditions and measurement directions. In the no-motion protocol, activation of the algorithm introduced slight unsharpness and increased heterogeneity in the high-frequency domain. Patient motion degrades CBCT spatial resolution. The proprietary correction algorithm may provide partial improvement under selected conditions, particularly at lower motion magnitudes, but it does not replace the need for careful patient stabilization during image acquisition and should be used cautiously when motion is unlikely.
Ahmed Z Abdelkarim, Ahmed Abdou, Scott Lozanoff, Nader Nabil Fouad Rezallah, Taraneh Maghsoodi-Zahedi, Sonam Khurana. Effect of a proprietary motion artifact correction algorithm on CBCT spatial resolution: an in vitro phantom study.. Oral surgery, oral medicine, oral pathology and oral radiology. 2026; doi:10.1016/j.oooo.2026.06.021.
Ahmed Z Abdelkarim, Ahmed Abdou, Scott Lozanoff, Nader Nabil Fouad Rezallah, Taraneh Maghsoodi-Zahedi, Sonam Khurana (2026). Effect of a proprietary motion artifact correction algorithm on CBCT spatial resolution: an in vitro phantom study.. Oral surgery, oral medicine, oral pathology and oral radiology. https://doi.org/10.1016/j.oooo.2026.06.021
Ahmed Z Abdelkarim, Ahmed Abdou, Scott Lozanoff, Nader Nabil Fouad Rezallah, Taraneh Maghsoodi-Zahedi, Sonam Khurana. Effect of a proprietary motion artifact correction algorithm on CBCT spatial resolution: an in vitro phantom study.[J]. Oral surgery, oral medicine, oral pathology and oral radiology, 2026 doi:10.1016/j.oooo.2026.06.021.
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出