基于实时人工智能的流动导向微导管尖端标记检测

Real-Time Artificial Intelligence-Based Detection of Flow-Directed Microcatheter Tip Markers.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

血流导向微导管(FDMs)具有高度的灵活性和跟踪性,可用于进入远端动脉供血分支。然而,其不透射线尖端标记较小,在透视下难以识别。本研究评估了一种术中基于人工智能(AI)的系统,用于实时检测FDM尖端标记。我们回顾性分析了连续10例使用该AI系统进行脑膜中动脉栓塞治疗慢性硬膜下血肿的病例。在微导管置入过程中,以逐帧方式评估检测率。根据导管直径(1.5 Fr vs. 1.3 Fr)以及导航过程中是微导管尖端还是导丝尖端先行推进,进行了探索性亚组分析。共分析25个FDM置入场景。平均微导管导航时间为2.1分钟。该系统的精确率、召回率和检测率分别为95%、51%和50%;检测率计算为所有分析帧中真阳性帧的比例。尽管这是探索性亚组分析,但1.5-Fr微导管的检测率似乎高于1.3-Fr微导管(78% vs. 46%;p = 0.006)。此外,当微导管尖端先于导丝尖端推进时,检测率显著高于相反情况(65% vs. 43%;p = 0.002)。尽管检测率为50%,但该AI系统可能有助于术者在透视下难以识别尖端位置时进行辨认。需要更大规模的多中心研究来验证这些发现。

Abstract

Flow-directed microcatheters (FDMs) offer high flexibility and trackability for accessing distal arterial feeders. However, the radiopaque tip marker is small, making it difficult to identify on fluoroscopy. This study evaluated an intraoperative artificial intelligence (AI)-based system for real-time detection of FDM tip markers. We retrospectively analyzed 10 consecutive cases of middle meningeal artery embolization for chronic subdural hematoma using the AI-based system. The detection rate was evaluated on a frame-by-frame basis during microcatheter placement. Exploratory subgroup analyses were performed based on the catheter diameter (1.5 Fr vs. 1.3 Fr) and whether the microcatheter tip or guidewire tip advanced ahead during navigation. Twenty-five FDM placement scenes were analyzed. Mean microcatheter navigation time was 2.1 min. The precision, recall, and detection rate of the system were 95%, 51%, and 50%, respectively; the detection rate was calculated as the proportion of true-positive frames among all analyzed frames. Although this was an exploratory subgroup analysis, the detection rate appeared higher for the 1.5-Fr microcatheters than that for the 1.3-Fr microcatheters (78% vs. 46%; p = 0.006). Furthermore, the detection rate was significantly higher when the microcatheter tip advanced ahead of the guidewire tip than vice versa (65% vs. 43%; p = 0.002). Although the detection rate was 50%, the AI system may help operators recognize the tip position when it becomes difficult to identify on fluoroscopy. Larger multicenter studies are required to validate these findings.

如何引用

AMA

Yuya Sakakura, Kenichi Kono, Takeshi Fujimoto. Real-Time Artificial Intelligence-Based Detection of Flow-Directed Microcatheter Tip Markers.. Journal of neuroendovascular therapy. 2026; doi:10.5797/jnet.oa.2026-0100.

APA

Yuya Sakakura, Kenichi Kono, Takeshi Fujimoto (2026). Real-Time Artificial Intelligence-Based Detection of Flow-Directed Microcatheter Tip Markers.. Journal of neuroendovascular therapy. https://doi.org/10.5797/jnet.oa.2026-0100

GB/T 7714

Yuya Sakakura, Kenichi Kono, Takeshi Fujimoto. Real-Time Artificial Intelligence-Based Detection of Flow-Directed Microcatheter Tip Markers.[J]. Journal of neuroendovascular therapy, 2026 doi:10.5797/jnet.oa.2026-0100.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI