兔(Oryctolagus cuniculi)股骨在置入皮质螺钉、锁定螺钉或克氏针后的抗扭强度降低无显著差异

Torsional Strength Reduction of Rabbit (Oryctolagus cuniculi) Femora is Not Significantly Different after Insertion of Cortical Screws, Locking Screws or Kirschner Wires.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

本研究旨在量化兔股骨中插入三种常见骨科植入物(皮质螺钉、锁定螺钉和克氏针)后植入物引起的骨丢失及扭转力学特性的变化。研究使用18根骨骼成熟的新西兰白兔股骨。股骨被随机分配接受由自攻皮质螺钉、自攻锁定螺钉或光滑克氏针创建的雙皮质缺损(每组6根)。植入物插入和取出后,标本接受微型计算机断层扫描以量化归一化的植入物周围骨体积丢失,随后进行扭转力学测试至失败。定性评估显示,与克氏针相比,两种螺钉类型在插入部位的骨碎片更大。微型计算机断层扫描分析显示,皮质螺钉插入后总骨体积丢失(11.7±3.4%)显著大于锁定螺钉(8.6±1.4%;=0.0459)。在生物力学测试中,各组间在扭转刚度、屈服扭矩、极限旋转和极限扭矩方面未观察到差异。这些发现表明,尽管自攻皮质螺钉在兔股骨中引起的体积骨丢失大于锁定螺钉,但这种损伤并未转化为所测试植入物之间在扭转刚度、屈服扭矩、极限旋转或极限扭矩方面的显著差异。因此,仅植入物插入本身不太可能是兔特定骨折修复技术中所见灾难性失败的主要因素。

Abstract

This study aims to quantify implant-induced bone loss and changes in torsional mechanical properties following insertion of three common orthopaedic implants - cortical screws, locking screws and Kirschner wires in rabbit femora. Eighteen femora from skeletally mature New Zealand White rabbits were used. Femora were randomly assigned to receive a bicortical defect created using a self-tapping cortical screw, self-tapping locking screw or smooth Kirschner wire ( = 6 per group). Following implant insertion and removal, specimens underwent micro-computed tomography to quantify normalized peri-implant bone volume loss, followed by torsional mechanical testing to failure. Qualitative assessment revealed larger bone chips at the insertion site for both screw types compared with Kirschner wires. Micro-computed tomography analysis demonstrated significantly greater total bone volume loss following cortical screw insertion (11.7 ± 3.4%) compared with locking screws (8.6 ± 1.4%; = 0.0459). No differences were observed between groups in torsional rigidity, yield torque, ultimate rotation and ultimate torque during biomechanical testing. These findings indicate that, although self-tapping cortical screws induce greater volumetric bone loss than locking screws in rabbit femora, this damage does not translate into significant differences in torsional rigidity, yield torque, ultimate rotation or ultimate torque among the tested implants. Implant insertion alone is therefore unlikely to be the primary contributor to catastrophic failure seen with specific fracture repair techniques in the rabbit.

如何引用

AMA

Anna M Massie, Karl E Hoenecke, Chet Friday, Badhan Saha, Farzad Abdi, Andrew Cagliostro, et al. Torsional Strength Reduction of Rabbit (Oryctolagus cuniculi) Femora is Not Significantly Different after Insertion of Cortical Screws, Locking Screws or Kirschner Wires.. Veterinary and comparative orthopaedics and traumatology : V.C.O.T. 2026; doi:10.1055/a-2934-5166.

APA

Anna M Massie, Karl E Hoenecke, Chet Friday, Badhan Saha, Farzad Abdi, Andrew Cagliostro, et al (2026). Torsional Strength Reduction of Rabbit (Oryctolagus cuniculi) Femora is Not Significantly Different after Insertion of Cortical Screws, Locking Screws or Kirschner Wires.. Veterinary and comparative orthopaedics and traumatology : V.C.O.T. https://doi.org/10.1055/a-2934-5166

GB/T 7714

Anna M Massie, Karl E Hoenecke, Chet Friday, Badhan Saha, Farzad Abdi, Andrew Cagliostro, et al. Torsional Strength Reduction of Rabbit (Oryctolagus cuniculi) Femora is Not Significantly Different after Insertion of Cortical Screws, Locking Screws or Kirschner Wires.[J]. Veterinary and comparative orthopaedics and traumatology : V.C.O.T, 2026 doi:10.1055/a-2934-5166.