基于机器学习的临床显著EDTA依赖性假性血小板减少症样本的决策逻辑
Decision-making logic for samples of clinically significant EDTA-dependent pseudothrombocytopenia with machine learning.
Decision-making logic for samples of clinically significant EDTA-dependent pseudothrombocytopenia with machine learning.
为建立实用且具有临床意义的判定标准,用于识别疑似样本中具有显著意义的血小板聚集,该标准以血小板计数偏差超过参考变化值(RCV)为定义。收集疑似聚集样本。枸橼酸盐/EDTA抗凝静脉血在5、30、60和120分钟时间点通过阻抗法(PLT-I)进行血小板计数及血涂片分析。人工显微镜检查(30分钟涂片)评估聚集簇大小/数量,而数字细胞形态学(PLT-pro)对所有时间点的聚集簇进行定量。以5分钟枸橼酸盐血小板计数为参考,计算EDTA PLT-I的相对偏差(d%)。临床显著聚集阳性样本定义为d%超过阴性RCV(PLT < 100 × 10⁹/L时为30.02%;PLT ≥ 100 × 10⁹/L时为20.92%)。样本随机分为训练集(70%)和测试集(30%),以建立逻辑回归和决策树模型。共86例患者贡献338个时间点结果;183个归类为阳性,155个为阴性。人工显微镜检查显示,计数含≥5个血小板的聚集簇(相较于≥3个)提高了特异性(p = 0.083)。≥3个聚集簇、每簇含≥5个血小板的判定标准达到100%灵敏度和33.3%特异性。通过Boruta算法,PLT-I和PLT-pro为关键判别指标。在测试集中,决策树模型优于逻辑回归(灵敏度:92.42% vs. 89.39%;特异性:94.29% vs. 94.29%;准确率:93.07% vs. 91.09%)。建立了人工显微镜检查标准(≥3个含≥5个血小板的聚集簇)和数字分析标准(PLT-pro 27-97且PLT-I < 58 × 10⁹/L,或PLT-pro ≥97且PLT-I < 164 × 10⁹/L)。然而,这些阈值仅为探索性结果,在用于临床决策之前必须在本实验室进行独立验证。
To establish practical, clinically meaningful criteria for identifying significant platelet aggregation in suspected samples, defined by a platelet-count deviation exceeding the reference change value (RCV). Suspected aggregated samples were collected. Venous blood anticoagulated with citrate/EDTA underwent platelet counting via impedance method (PLT-I) and blood smear analysis at 5-, 30-, 60-, and 120-min intervals. Manual microscopy (30-min smears) assessed aggregation cluster size/number, while digital cytomorphology (PLT-pro) quantified clusters at all time points. Using 5-min citrate platelet counts as reference, relative deviations (d%) of EDTA PLT-I were calculated. Clinically significant aggregation-positive samples were defined as d% exceeding the negative RCV (30.02% for PLT < 100 × 10(Zhang et al., 2025/L; 20.92% for PLT ≥ 100 × 10(Zhang et al., 2025/L). Samples were randomly split into training (70%) and test (30%) sets to develop logistic regression and decision-tree models. A total of 86 patients contributed 338 time-point results; 183 were classified as positive and 155 as negative. Manual microscopy showed that counting clusters with ≥5 platelets (vs. ≥ 3) improved specificity (p = 0.083). A criterion of ≥3 clusters, each containing ≥5 platelets, achieved 100% sensitivity and 33.3% specificity. PLT-I and PLT-pro were key discriminators via the Boruta algorithm. In the test set, the decision-tree model outperformed logistic regression (sensitivity: 92.42% vs. 89.39%; specificity: 94.29% vs. 94.29%; accuracy: 93.07% vs. 91.09%). Criteria were established for manual microscopy (≥ 3 clusters with ≥5 platelets) and digital analysis (PLT-pro 27-97 with PLT-I < 58 × 10(Zhang et al., 2025/L, or PLT-pro ≥97 with PLT-I < 164 × 10(Zhang et al., 2025/L). However, these thresholds are merely exploratory results, and independent validation must be performed in this laboratory before they can be used for clinical decision-making.
Xin Meng, Na Li, Liu Yang, Xiaoqin Wang, Wenjing Wu. Decision-making logic for samples of clinically significant EDTA-dependent pseudothrombocytopenia with machine learning.. Clinica chimica acta; international journal of clinical chemistry. 2026; doi:10.1016/j.cca.2026.121291.
Xin Meng, Na Li, Liu Yang, Xiaoqin Wang, Wenjing Wu (2026). Decision-making logic for samples of clinically significant EDTA-dependent pseudothrombocytopenia with machine learning.. Clinica chimica acta; international journal of clinical chemistry. https://doi.org/10.1016/j.cca.2026.121291
Xin Meng, Na Li, Liu Yang, Xiaoqin Wang, Wenjing Wu. Decision-making logic for samples of clinically significant EDTA-dependent pseudothrombocytopenia with machine learning.[J]. Clinica chimica acta; international journal of clinical chemistry, 2026 doi:10.1016/j.cca.2026.121291.