同一样本,不同结果:英国四家实验室氯氮平检测测量的差异

Same sample, different result: Variation in clozapine assay measurement in four UK laboratories.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

氯氮平血药浓度的实验室检测对于安全有效的治疗至关重要,且通常被认为在不同检测机构间具有准确性和可重复性。色谱法检测需要批次特异性设置,并使用不同的设备和技术进行操作,这可能引入误差或变异性。然而,对检测机构间一致性的系统评价尚有限。旨在评估英国四家提供常规治疗药物监测的实验室对氯氮平和去甲氯氮平血药浓度测定的分析准确性和可重复性。基于实验室的分析准确性和重复性研究。将一组15份加标血浆样本分发至英国四家提供常规治疗药物监测的实验室,样本覆盖氯氮平和去甲氯氮平的临床相关浓度范围(74-1443 ng/mL)。各实验室在两次独立场合对相同样本进行分析。使用回归和一致性分析评估相对于已知值的分析准确性及实验室内一致性。所有实验室均显示测定浓度与已知浓度之间呈强线性关系(r = 0.932-0.999)。实验室A和C在兩轮检测中斜率接近1(0.94-0.99),平均偏差极小(-28至+22 ng/mL),氯氮平(-89至+51 ng/mL)和去甲氯氮平(-71至+71 ng/mL)的一致性界限较窄。相比之下,实验室B和D在重复检测中一致性降低,第二轮中斜率增大、偏差为正且一致性界限增宽。实验室B的氯氮平斜率从0.99增至1.33,一致性界限增宽至-320至+673 ng/mL。实验室D偏差更大,氯氮平斜率从1.18增至1.59,平均偏差升至+349 ng/mL,一致性界限极宽(-354至+1053 ng/mL)。使用LC-MS和HPLC进行的氯氮平实验室检测并非始终准确或随时间可重复。部分实验室表现出显著偏差,报告的测定浓度远高于真实值。

Abstract

Laboratory measurement of clozapine blood concentrations is central to safe and effective treatment and is commonly assumed to be accurate and reproducible across providers. Chromatographic assays require batch-specific setup and are performed using different equipment and techniques, which may introduce error or variability. However, systematic evaluation of inter-provider consistency is limited. To assess the analytical accuracy and reproducibility of clozapine and norclozapine blood concentration measurements across four UK laboratories providing routine therapeutic drug monitoring. Laboratory-based analytical accuracy and repeatability study. A panel of 15 spiked plasma samples covering a clinically relevant concentration range for clozapine and norclozapine (74-1443 ng/mL) was distributed to four UK laboratories providing routine therapeutic drug monitoring. Each laboratory analysed the same samples on two separate occasions. Analytical accuracy relative to known values and intra-laboratory consistency were assessed using regression and agreement analyses. All laboratories demonstrated strong linearity between measured and known concentrations (r = 0.932-0.999). Laboratories A and C showed slopes close to unity (0.94-0.99), minimal mean bias (-28 to +22 ng/mL), and narrow limits of agreement for clozapine (-89 to +51 ng/mL) and norclozapine (-71 to +71 ng/mL) across both rounds. In contrast, laboratories B and D showed reduced agreement on repeat testing, with increased slopes, positive bias, and widened limits of agreement in the second round. For laboratory B, the clozapine slope increased from 0.99 to 1.33, with limits of agreement widening to -320 to +673 ng/mL. Laboratory D showed greater deviation, with the clozapine slope increasing from 1.18 to 1.59 and mean bias rising to +349 ng/mL, with very wide limits of agreement (-354 to +1053 ng/mL). Laboratory assays of clozapine using LC-MS and HPLC are not always accurate or reproducible over time. Some laboratories show a significant bias, reporting measured concentrations much higher than the true value.

如何引用

AMA

Matthew Atkins, James Fritz, Mary Rose Hilaire, David Taylor. Same sample, different result: Variation in clozapine assay measurement in four UK laboratories.. Therapeutic advances in psychopharmacology. 2026; doi:10.1177/20451253261473111.

APA

Matthew Atkins, James Fritz, Mary Rose Hilaire, David Taylor (2026). Same sample, different result: Variation in clozapine assay measurement in four UK laboratories.. Therapeutic advances in psychopharmacology. https://doi.org/10.1177/20451253261473111

GB/T 7714

Matthew Atkins, James Fritz, Mary Rose Hilaire, David Taylor. Same sample, different result: Variation in clozapine assay measurement in four UK laboratories.[J]. Therapeutic advances in psychopharmacology, 2026 doi:10.1177/20451253261473111.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI

Therapeutic advances in psychopharmacology 的其他文献