质子治疗多环双散射系统的建模与优化
Modeling and optimization of a multiple-ring double scattering system for proton therapy.
Modeling and optimization of a multiple-ring double scattering system for proton therapy.
本研究探讨质子束散射系统,旨在提高束流整形效率,并在强度受限条件下增加低能束流在辐射野内的传输。这些改进对于眼质子治疗(OPT)中高剂量率辐射野的输送,乃至潜在FLASH放疗应用中超高剂量率的输送尤为重要。研究开发了一套系统方法,用于设计多环双散射系统,在控制各环出射动能的同时,在射野均匀性约束下最大化传输效率。该框架虽具有普遍适用性,但其具体动机源于亥姆霍兹柏林材料与能源研究中心(HZB)使用70 MeV回旋加速器开展质子眼治疗的临床与实验需求。研究实现了基于Python的图形用户界面(GUI),集成了针对2至4个同心环的数学优化模型,在均匀性约束下最大化传输。该软件整合了能量降级、射程移位以及将优化散射角转换为物理尺寸的功能。数学优化准确复现了Takada(1994)报道的双环结果,并将方法扩展至三环和四环配置。四环系统实现了53±2%的传输率,剂量均匀性为±3%,对应于30 mm射野尺寸下100 nA时77 Gy/s的剂量率,超过了FLASH条件的阈值。布拉格峰测量和使用TOPAS进行的蒙特卡罗模拟显示,远端半影低于1 mm,水中射程与预测值的偏差在0.1 mm以内,支持了所采用的能量降级方法。所开发的GUI为设计和优化双散射系统提供了实用工具,与单散射方法相比显著提高了传输效率。四环设计使束流传输比单散射提高了十倍以上,使其成为小野高剂量率照射的有前景的候选方案。
This study investigates proton beam scattering systems with the aim of improving beam-shaping efficiency and increasing transmission within the radiation field for low-energy beams under intensity-limited conditions. These improvements are particularly relevant for delivering radiation fields at high dose rates for ocular proton therapy (OPT) and even at ultra-high dose rates for potential FLASH radiotherapy applications. A systematic methodology was developed to design multiple-ring double-scattering systems that maximize transmission under a field-homogeneity constraint while controlling the outgoing kinetic energy after each ring. Although generally applicable, the framework was specifically motivated by clinical and experimental requirements for proton-based ocular therapy using a 70 MeV cyclotron at the Helmholtz-Zentrum Berlin für Materialien und Energie (HZB). A Python-based graphical user interface (GUI) was implemented to integrate a mathematical optimization model for two to four concentric rings that maximizes transmission under homogeneity constraints. The software incorporates energy degradation, range shifting, and the conversion of optimized scattering angles into physical dimensions. The mathematical optimization accurately reproduced the two-ring results reported by Takada (1994) and extended the approach to three- and four-ring configurations. The four-ring system achieved a transmission of 53±2% with a dose homogeneity of ±3%, corresponding to a dose rate of 77Gy/s at 100nA for a 30 mm field size, exceeding the threshold for FLASH conditions. Bragg-peak measurements and Monte Carlo simulations using TOPAS showed a distal penumbra below 1 mm and a water range agreement within 0.1 mm of predictions, supporting the adopted energy degradation methodology. The developed GUI provides a practical tool for designing and refining double-scattering systems, significantly improving transmission compared with single-scattering approaches. The four-ring design increased beam transmission more than tenfold compared with single scattering, making it a promising candidate for high-dose-rate irradiation in small fields.
J P de Oliveira Lima, A Weber, G Kourkafas, A Denker, J Heufelder. Modeling and optimization of a multiple-ring double scattering system for proton therapy.. Zeitschrift fur medizinische Physik. 2026; doi:10.1016/j.zemedi.2026.07.006.
J P de Oliveira Lima, A Weber, G Kourkafas, A Denker, J Heufelder (2026). Modeling and optimization of a multiple-ring double scattering system for proton therapy.. Zeitschrift fur medizinische Physik. https://doi.org/10.1016/j.zemedi.2026.07.006
J P de Oliveira Lima, A Weber, G Kourkafas, A Denker, J Heufelder. Modeling and optimization of a multiple-ring double scattering system for proton therapy.[J]. Zeitschrift fur medizinische Physik, 2026 doi:10.1016/j.zemedi.2026.07.006.
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出