{"_":"瑞马匹拉嗪(一种用于即时评估肾小球滤过率的荧光示踪剂)对住院患者常进行的临床诊断检测的干扰缺失。","i":["体外"]}

{"_":"Absence of Interference on Clinical Diagnostic Assays Often Performed on Hospitalized Patients by Relmapirazin, a Fluorescent Tracer Agent for Point-of-Care Assessment of Glomerular Filtration Rate.","i":["in vitro"]}

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

FDA近期批准了一种医疗器械,该设备通过经皮检测荧光示踪剂relmapirazin,在即时检验中评估肾小球滤过率。此前针对该示踪剂的体外和体内研究显示,其安全性/毒理学担忧可忽略不计,因而得以推进至I期、II期和III期人体临床研究。既然该示踪剂已进入临床应用,本文研究了该示踪剂可能对入院及住院患者常规临床诊断检测产生的潜在干扰。检测方法遵循CLSI EP07-A3《临床化学干扰试验》指南进行干扰测定。在所评估的代谢和心脏检测组合中,未检测到relmapirazin的干扰。

Abstract

The FDA has recently approved a medical device which determines a glomerular filtration rate assessment at the point-of-care by transdermal detection of the fluorescent tracer agent relmapirazin. Previous and studies on this agent yielded negligible safety/toxicology concern resulting in advancement to phase I, II, and III human clinical studies. Now that clinical use is occurring, herein we investigated possible interference this agent may induce with typical clinical diagnostic assays given to incoming and hospitalized patients. The testing methodology followed the CLSI EP07-A3 Interference Testing in Clinical Chemistry guideline for determination of interference. No interference from relmapirazin was detected in the suite of metabolic and cardiac assays evaluated.

如何引用

AMA

Richard B Dorshow, Nancy Morrison, Margot Borgel. {"_":"Absence of Interference on Clinical Diagnostic Assays Often Performed on Hospitalized Patients by Relmapirazin, a Fluorescent Tracer Agent for Point-of-Care Assessment of Glomerular Filtration Rate.","i":["in vitro"]}. International journal of toxicology. 2026; doi:10.1177/10915818261477667.

APA

Richard B Dorshow, Nancy Morrison, Margot Borgel (2026). {"_":"Absence of Interference on Clinical Diagnostic Assays Often Performed on Hospitalized Patients by Relmapirazin, a Fluorescent Tracer Agent for Point-of-Care Assessment of Glomerular Filtration Rate.","i":["in vitro"]}. International journal of toxicology. https://doi.org/10.1177/10915818261477667

GB/T 7714

Richard B Dorshow, Nancy Morrison, Margot Borgel. {"_":"Absence of Interference on Clinical Diagnostic Assays Often Performed on Hospitalized Patients by Relmapirazin, a Fluorescent Tracer Agent for Point-of-Care Assessment of Glomerular Filtration Rate.","i":["in vitro"]}[J]. International journal of toxicology, 2026 doi:10.1177/10915818261477667.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI

International journal of toxicology 的其他文献