用于独立记录仪器基于事件同步的机械诱导基准标记系统

Mechanically-induced fiducial marker system for event-based synchronization of independent recording instruments.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

许多实验和临床测量需要从各自独立、不共享时钟的仪器同时采集数据。我们描述了一种通过基于事件的同步实现此类配对的装置:一台仪器发出的电触发信号驱动一个由Arduino控制的步进执行器,该执行器在另一台仪器所感测的位置产生短暂的机械事件,从而留下一个基准标记,用于事后进行时间对齐。其应用背景是将高分辨率测压(HRM)与会阴超声同步,以对压力性尿失禁进行表型分型。台架表征显示,装置延迟为(均值±标准差,),其中执行器的机电响应占平均延迟的,并且占试验间变异的主要部分(即的标准差)。在超声-HRM应用中,同步不确定度为(个标记),该不确定度由HRM而非装置决定,装置本身的抖动贡献可忽略不计。该方法可推广至这样的仪器配对:其中一台仪器提供与其记录时间线具有已知关系的稳定电触发,另一台为具有足够带宽的机械敏感接收器,前提是标记不会污染所关注的信号。

Abstract

Many experimental and clinical measurements require simultaneous acquisition from independent instruments that do not share a clock. We describe a device that enables such pairings via event-based synchronization: an electrical trigger from one instrument drives an Arduino-controlled stepper actuator that produces a brief mechanical event at a location sensed by the other instrument, leaving a fiducial marker for post-hoc temporal alignment. The motivating application is synchronization of high-resolution manometry (HRM) with transperineal ultrasound for phenotyping stress urinary incontinence. Bench characterization yielded a device latency of (mean SD, ), of which the electromechanical response of the actuator accounts for of the mean latency and the dominant share of the trial-to-trial variation (i.e. the SD of ). For the ultrasound-HRM application, the synchronization uncertainty is ( markers), set by the HRM rather than the device, whose intrinsic jitter contributes negligibly. The approach generalizes to instrument pairings in which one instrument provides a stable electrical trigger with a known relation to its recording timeline and the other is a mechanically-sensitive receiver of adequate bandwidth, provided the marker does not contaminate the signal of interest.

如何引用

AMA

Kourosh Kalayeh, John O L DeLancey, James A Ashton-Miller, J Brian Fowlkes. Mechanically-induced fiducial marker system for event-based synchronization of independent recording instruments.. Measurement science & technology. 2026; doi:10.1088/1361-6501/ae92a4.

APA

Kourosh Kalayeh, John O L DeLancey, James A Ashton-Miller, J Brian Fowlkes (2026). Mechanically-induced fiducial marker system for event-based synchronization of independent recording instruments.. Measurement science & technology. https://doi.org/10.1088/1361-6501/ae92a4

GB/T 7714

Kourosh Kalayeh, John O L DeLancey, James A Ashton-Miller, J Brian Fowlkes. Mechanically-induced fiducial marker system for event-based synchronization of independent recording instruments.[J]. Measurement science & technology, 2026 doi:10.1088/1361-6501/ae92a4.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI