结合4D-CT成像与机械负荷分析的肘关节不稳诊断工具的开发

Development of a Diagnostic Tool for Elbow Instability Combining 4D-CT Imaging and Mechanical Load Analysis.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

在临床实践中,肘关节不稳定的诊断通常依赖于医生手动进行应力测试,可结合或不结合透视或超声检查,这可能导致施力大小不一和主观判读差异。在动态成像过程中采用标准化方法施加受控的机械负荷可能提高可重复性。本研究的目的是开发一种与CT兼容的诊断装置,用于装置引导下的肘关节不稳定评估,并同时进行力和位置测量,并开展初步的尸体可行性测试。技术开发研究,包含初步尸体可行性测试。在初始原型基础上,本文描述了后续原型MELBO的开发,其运动学参数经过调整、稳定性得到改善,同时涉及力传感器的验证,并呈现了首批尸体测试结果。MELBO诊断工具能够在4D-CT采集过程中实现装置引导下的受控肘关节运动,并允许同时记录力和位置数据。所获取的动态CT数据集适用于肘关节稳定性的定性评估。未观察到相关的装置相关成像伪影。本研究证明了使用MELBO诊断工具在4D-CT过程中进行装置引导下同步力和位置测量的技术可行性。这些发现为进一步的尸体研究和未来旨在标准化负荷条件下定量评估肘关节不稳定的临床研究提供了基础。

Abstract

Diagnosis of elbow instability in clinical practice commonly relies on manual stress testing by the physician with or without combination of fluoroscopy or ultrasound, which may result in variable force application and subjective interpretation. A standardized method for applying controlled mechanical loading during dynamic imaging may improve reproducibility. The objective of this study was to develop and perform initial cadaveric feasibility testing of a CT-compatible diagnostic device for device-guided assessment of elbow instability with simultaneous force and positional measurement. Technical development study with initial cadaveric feasibility testing. Building upon the initial prototype the present paper, describes the development of MELBO the follow-up prototype with adjusted kinematics and improved stability, the validation of the force sensor is addressed, and the first cadaver test results are presented. The MELBO diagnostic tool enabled a controlled device-guided elbow motion during 4D-CT acquisition and allowed simultaneous recording of force and positional data. The acquired dynamic CT datasets were suitable for qualitative assessment of elbow joint stability. No relevant device-related imaging artifacts were observed. The present study demonstrates the technical feasibility of device-guided simultaneous force and positional measurement during 4D-CT using the MELBO diagnostic tool. These findings provide the basis for further cadaveric studies and future clinical investigations aimed at quantitative assessment of elbow instability under standardized loading conditions.

如何引用

AMA

Jonas Gehring, Fabian Forster, Markus Meindl, Martin Maerz, Michael Simon, Nina Renner, et al. Development of a Diagnostic Tool for Elbow Instability Combining 4D-CT Imaging and Mechanical Load Analysis.. Biomedical engineering and computational biology. 2026; doi:10.1177/11795972261472224.

APA

Jonas Gehring, Fabian Forster, Markus Meindl, Martin Maerz, Michael Simon, Nina Renner, et al (2026). Development of a Diagnostic Tool for Elbow Instability Combining 4D-CT Imaging and Mechanical Load Analysis.. Biomedical engineering and computational biology. https://doi.org/10.1177/11795972261472224

GB/T 7714

Jonas Gehring, Fabian Forster, Markus Meindl, Martin Maerz, Michael Simon, Nina Renner, et al. Development of a Diagnostic Tool for Elbow Instability Combining 4D-CT Imaging and Mechanical Load Analysis.[J]. Biomedical engineering and computational biology, 2026 doi:10.1177/11795972261472224.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI