[不同类型牙科植入物的扭矩值和主要稳定性取决于钻头设计]

[The values of torque and primary stability of different types of dental implants depending on drill design].

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

探讨在低密度骨中,采用标准螺旋钻与阶梯钻预备种植床时,非侵袭性螺纹与“活性”螺纹种植体的扭矩值及种植体稳定系数(ISQ)的表现。实验模型采用猪胸骨。共使用40枚种植体,根据螺纹设计分为两组,每组各含两种类型种植体10枚。第一组采用螺旋钻进行骨床预备,第二组采用阶梯钻预备。使用手术马达及测力扳手测量峰值扭矩,并采用Osstell设备配合Smartpeg和Multipeg pins评估种植体稳定系数。第一组中,使用任何类型种植体均无显著差异。第二组中,使用阶梯钻植入“活性”螺纹种植体时,所研究的参数显著增加。扭矩值增加1.7倍。Smartpeg和Multipeg pins的稳定系数平均分别增加8%和7%。当比较采用标准钻植入的非侵袭性螺纹圆柱形种植体与采用阶梯钻植入的“活性”螺纹锥形种植体时,最大扭矩及稳定系数的增加分别为80%、8%和7%。在低密度骨中操作时,无论使用螺旋钻还是阶梯钻,采用非侵袭性螺纹圆柱形种植体均表现为较低的扭矩值和稳定系数。在低密度骨中,采用阶梯钻植入“活性”螺纹锥形种植体是有效的。

Abstract

To investigate the values of torque and implant stability quotient of implants with non-aggressive and «active» threads placed using standard spiral and stepped drills. As the experimental model was used pig's breastbone. Forty implants were divided in two groups which included 10 implants of each type according to thread design. In the first group bone site preparation was conducted with spiral drills, in the second one - using stepped drill. Peek torque values were measured with surgical motor and dynamometric key and to assess implant stability quotient was used Osstell device with Smartpeg and Multipeg pins. In the first group, there was no significant difference using any type of implant. In the second group, there was a significant increase in the studied parameters during the use of implants with «active» threads with a stepped drill. Torque values increased in 1.7 times. The increase in stability quotient for both Smartpeg and Multipeg pins in average was 8% and 7% respectively. When comparing cylindrical implants with non-aggressive threads with standard drills and conical implants with «active» threads with a stepped drill, the increase of maximum torque and stability quotient was 80%, 8% and 7% respectively. The use of cylindrical implants with non-aggressive threads when working in low density bone both when using spiral or stepped drills, is characterized by low values of torque and stability quotient. In low density bone the placement of conical implants with «active» threads using a stepped drill is effective.

如何引用

AMA

B A Kudzaev, V A Badalyan, E M Chernovol. [The values of torque and primary stability of different types of dental implants depending on drill design].. Stomatologiia. 2026; doi:10.17116/stomat202610504111.

APA

B A Kudzaev, V A Badalyan, E M Chernovol (2026). [The values of torque and primary stability of different types of dental implants depending on drill design].. Stomatologiia. https://doi.org/10.17116/stomat202610504111

GB/T 7714

B A Kudzaev, V A Badalyan, E M Chernovol. [The values of torque and primary stability of different types of dental implants depending on drill design].[J]. Stomatologiia, 2026 doi:10.17116/stomat202610504111.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI

Stomatologiia 的其他文献