光子计数探测器4D CT用于原发性甲状旁腺功能亢进症的术前定位

Photon-Counting Detector 4D CT for Preoperative Localization in Primary Hyperparathyroidism.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

异常甲状旁腺组织的术前准确定位对于微创甲状旁腺切除术的成功至关重要。常规超声和Tc-99m甲氧基异丁基异腈闪烁显像存在局限性,尤其在异位病变中。本研究评估了光子计数探测器CT(PCD-CT)在原发性甲状旁腺功能亢进症患者中进行4D甲状旁腺成像的性能。在这项经IRB批准的双向性研究中,36例原发性甲状旁腺功能亢进症患者接受了术前4D PCD-CT检查,随后行甲状旁腺切除术。分别有34例和33例患者具备超声和甲氧基异丁基异腈闪烁显像资料。三名经过专科培训的盲法放射科医师独立审阅影像学检查,每位阅片者仅负责单一模态。影像学表现与手术及组织病理学结果(作为参考标准)进行比较。患者水平的定位率以双侧95%精确二项置信区间汇总;各模态间的比较为描述性。PCD-CT在36/36例患者中正确定位了异常甲状旁腺组织(100%;95% CI,90.3%-100%)。超声显示经手术证实的明确病变为12/34例(35%;95% CI,19.7%-53.5%),当纳入经证实的可疑表现后增至16/34例(47%;95% CI,29.8%-64.9%)。甲氧基异丁基异腈闪烁显像显示经手术证实的明确病变为19/33例(57%;95% CI,39.2%-74.5%);当纳入可疑表现后,经证实的定位增至23/33例(70%;95% CI,51.3%-84.4%)。PCD-CT定位了8例超声或闪烁显像未识别的病变,其中大多数为下极或异位病变。全部7例异位腺瘤均由PCD-CT定位,而超声定位3/7例,闪烁显像定位4/7例。光子计数4D CT在原发性甲状旁腺功能亢进症患者异常甲状旁腺的术前定位中显示出令人鼓舞的高定位率,尤其在异位腺瘤方面。

Abstract

Accurate preoperative localization of abnormal parathyroid tissue is essential for successful minimally invasive parathyroidectomy. Conventional imaging with ultrasound and Tc-99m sestamibi scintigraphy can be limited, particularly in ectopic disease. This study evaluated the performance of photon-counting detector CT (PCD-CT) for 4D parathyroid imaging in patients with primary hyperparathyroidism. In this IRB-approved ambispective study, 36 patients with primary hyperparathyroidism underwent preoperative 4D PCD-CT followed by parathyroidectomy. Ultrasound and sestamibi scintigraphy were available in 34 and 33 patients, respectively. Three blinded fellowship-trained radiologists independently reviewed imaging studies, with each reader assigned to a single modality. Imaging findings were compared with operative and histopathologic results, which served as the reference standard. Patient-level localization rates were summarized with two-sided 95% exact binomial confidence intervals; comparisons among modalities were descriptive. PCD-CT correctly localized abnormal parathyroid tissue in 36/36 patients (100%; 95% CI, 90.3%-100%). Ultrasound demonstrated surgically confirmed definite lesions in 12/34 patients (35%; 95% CI, 19.7%-53.5%), increasing to 16/34 (47%; 95% CI, 29.8%-64.9%) when confirmed equivocal findings were included. Sestamibi scintigraphy demonstrated definite surgically confirmed lesions in 19/33 patients (57%; 95% CI, 39.2%-74.5%); confirmed localization increased to 23/33 (70%; 95% CI, 51.3%-84.4%) when equivocal findings were included. PCD-CT localized lesions in 8 patients not identified by ultrasound or scintigraphy, most of which were inferior or ectopic. All 7 ectopic adenomas were localized by PCD-CT, compared with 3/7 by ultrasound and 4/7 by scintigraphy. Photon-counting 4D CT demonstrated promising high localization rate for preoperative localization of abnormal parathyroid glands in primary hyperparathyroidism, particularly with ectopic adenomas.

如何引用

AMA

Sean J Huls, Brian J Burkett, Lifeng Yu, Joel G Fletcher, Cynthia H McCollough, Nikkole M Weber, et al. Photon-Counting Detector 4D CT for Preoperative Localization in Primary Hyperparathyroidism.. AJNR. American journal of neuroradiology. 2026; doi:10.3174/ajnr.A9583.

APA

Sean J Huls, Brian J Burkett, Lifeng Yu, Joel G Fletcher, Cynthia H McCollough, Nikkole M Weber, et al (2026). Photon-Counting Detector 4D CT for Preoperative Localization in Primary Hyperparathyroidism.. AJNR. American journal of neuroradiology. https://doi.org/10.3174/ajnr.A9583

GB/T 7714

Sean J Huls, Brian J Burkett, Lifeng Yu, Joel G Fletcher, Cynthia H McCollough, Nikkole M Weber, et al. Photon-Counting Detector 4D CT for Preoperative Localization in Primary Hyperparathyroidism.[J]. AJNR. American journal of neuroradiology, 2026 doi:10.3174/ajnr.A9583.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI

AJNR. American journal of neuroradiology 的其他文献