体外生命支持中所用泵的压力-流量测量

Pressure - flow measurements of pumps used in extracorporeal life support.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

引言血液泵是体外生命支持(ECLS)回路的核心组件。目前主要使用现代离心泵和斜流泵,而非滚压泵。与滚压泵相比,离心泵和斜流泵具有预充量更低、溶血更少以及可实现回路小型化的优势。其缺点在于开放式设计会促进内部再循环,从而影响性能,并且在少数低流量研究中可能增加凝血激活和溶血的风险。本研究旨在提供现有泵的性能数据。方法研究人员联系制造商借用ECLS控制台并捐赠泵样本用于研究,不接受任何财务支持。缺失的泵由研究者自行购买。在模拟回路中,测量体积流量和压力。按照制造商标称的工作流量范围,在不同前负荷和后负荷水平下对泵进行评估。结果共测试了八种不同的ECLS泵,其中两种适用于儿科,六种适用于成人。六种为离心泵,两种为斜流设计。所有泵均为磁驱动,其中三种为磁悬浮。五种泵的叶轮为带罩(覆盖)设计。使用水或水-葡萄糖溶液作为测试流体,并就此讨论了介质的选择。为每种泵在其工作范围内生成了流量-压力图谱。经与制造商协商,未对泵进行直接比较。结论本研究报告了专为儿童和成人ECLS使用设计的血液泵的性能曲线,并讨论了设计特征的重要方面,以支持临床医生。

Abstract

IntroductionThe blood pump is a central component of the extracorporeal life support (ECLS) circuit. Modern centrifugal and diagonal pumps are mainly used in preference to roller pumps. Compared to roller pumps, centrifugal and diagonal pumps offer lower priming volume, less hemolysis and enable miniaturization of circuits. Drawback is the open design promoting internal recirculation, which influences performance and, in few studies at low flow, may increase risk of coagulation activation and hemolysis. The aim of this work was to provide performance data on available pumps.MethodsManufacturers were asked to lend ECLS consoles and donate pump samples for the study. No financial support was accepted. Missing pumps were bought by the investigators. In a mock-loop, measurements of volume flow and pressures were performed. Pumps were assessed over the operating flow range stated by the manufacturer at different preloads and afterload levels.ResultsEight different ECLS pumps were tested, two available for pediatric use and six for adult use. Six pumps were centrifugal and two were of diagonal design. All were magnetically driven and three were magnetically levitated. Five had impellers with shrouded (covered) design. Water or water-glucose were used as a test fluid and choice of medium is herewith also discussed. Flow-pressure maps for each pump were generated over its operating range. As agreed with manufacturers, no direct comparisons between pumps were made.ConclusionsPerformance curves for blood pumps designed for ECLS use in children and adults were reported and important aspects of design features discussed to support clinicians.

如何引用

AMA

Federico Rorro, Lisa Prahl Wittberg, Fabio Silvio Taccone, Nicholas A Barrett, Maximilain V Malfertheiner, Matteo Di Nardo, et al. Pressure - flow measurements of pumps used in extracorporeal life support.. Perfusion. 2026; doi:10.1177/02676591261476520.

APA

Federico Rorro, Lisa Prahl Wittberg, Fabio Silvio Taccone, Nicholas A Barrett, Maximilain V Malfertheiner, Matteo Di Nardo, et al (2026). Pressure - flow measurements of pumps used in extracorporeal life support.. Perfusion. https://doi.org/10.1177/02676591261476520

GB/T 7714

Federico Rorro, Lisa Prahl Wittberg, Fabio Silvio Taccone, Nicholas A Barrett, Maximilain V Malfertheiner, Matteo Di Nardo, et al. Pressure - flow measurements of pumps used in extracorporeal life support.[J]. Perfusion, 2026 doi:10.1177/02676591261476520.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI

Perfusion 的其他文献