协调M蛋白分析:一项国际性的、多中心的质谱再现性研究和电泳方法一致性评估
Harmonizing M-protein analysis: an international, multi-center reproducibility study of mass spectrometry and evaluation of concordance to electrophoretic methods.
Harmonizing M-protein analysis: an international, multi-center reproducibility study of mass spectrometry and evaluation of concordance to electrophoretic methods.
血清M蛋白分析是单克隆丙种球蛋白病评估的核心环节。定量和同种型鉴定通常分别通过血清蛋白电泳(SPE)或毛细管区带电泳(CZE)以及免疫固定电泳(IFE)或免疫分型(IT)来完成。尽管这些检测方法被广泛使用,但其敏感度有限且判读具有主观性。我们评估了一种自动化基质辅助激光解吸电离飞行时间质谱(MALDI-TOF MS)平台的跨站点重现性及其与电泳方法的一致性。我们制备了来自17例确诊单克隆蛋白患者的120份盲法、系列稀释血清标本,其中118份符合定量分析条件。五家学术实验室使用常规SPE/CZE、IFE/IT、Optilite总免疫球蛋白测定和MALDI-TOF MS(EXENT系统)对相同样本进行分析。我们评估了质量精密度、定量与定性一致性、跨站点重现性以及基于变异系数(CV)的变异性。质量精密度极佳,超过99%的结果在五站点均值的4以内。与IFE/IT的一致性为81.7%。Bland-Altman分析显示平均偏差范围为-1.14 g/L至4.61 g/L,并观察到同种型依赖性差异。SPE/CZE与MS的整体比较显示Passing-Bablok斜率中位数为1.004(IQR:0.055)。MS的跨站点重现性优于SPE/CZE,中位斜率分别为1.029(IQR:0.083)和0.997(IQR:0.113)。MS较SPE/CZE更常出现CV≤15%的结果(67.8% vs. 46.9%),尤其是在M蛋白浓度<10 g/L时。自动化MALDI-TOF MS表现出较强的跨站点重现性,并与电泳方法高度一致。残余变异性可归因于上游免疫球蛋白定量、IgM相关效应以及电泳整合的已知局限性。这些发现支持MS作为多中心环境下M蛋白分析的稳健且可 harmonize 的平台。
Serum M-protein analysis is central to the assessment of monoclonal gammopathies. Quantitation and isotyping are commonly performed by SPE or CZE and IFE or IT, respectively. While these assays are widely used, they are characterized by limited sensitivity and subjective interpretation. We evaluated the inter-site reproducibility of an automated MALDI-TOF mass spectrometry (MS) platform and its concordance with electrophoretic methods. We prepared 120 blinded, serially diluted serum specimens from 17 patients with confirmed monoclonal proteins, of which 118 were eligible for quantitative analyses. Five academic laboratories analyzed identical samples using routine SPE/CZE, IFE/IT, Optilite total immunoglobulin measurements and MALDI-TOF MS (the EXENT System). We assessed mass precision, quantitative and qualitative agreement, inter-site reproducibility, and CV-based variability. Mass precision was excellent with >99 % of results within 4 of the five-site mean. Agreement with IFE/IT was 81.7 %. Bland-Altman analysis resulted in mean biases ranging from -1.14 g/L to 4.61 g/L with isotype-dependent differences observed. Global comparison of SPE/CZE to MS resulted in a median Passing-Bablok slope of 1.004 (IQR: 0.055). Inter-site reproducibility was improved for MS over SPE/CZE with median slopes of 1.029 (IQR: 0.083) and 0.997 (IQR: 0.113), respectively. Results with CV ≤15 % were more frequent for MS than by SPE/CZE (67.8% vs. 46.9 %), particularly at MP concentrations <10 g/L. Automated MALDI-TOF MS demonstrated strong inter-site reproducibility and close concordance with electrophoretic methods. Residual variability was attributable to upstream immunoglobulin quantification, IgM-related effects and known limitations of electrophoretic integration. These findings support MS as a robust and harmonizable platform for M-protein analysis in multi-center settings.
Katie Thoren, Lauren Campbell, Rosa Pello, María Ángeles Iglesias, Raúl Mateos Pablos, Cristina Agulló, et al. Harmonizing M-protein analysis: an international, multi-center reproducibility study of mass spectrometry and evaluation of concordance to electrophoretic methods.. Clinical chemistry and laboratory medicine. 2026; doi:10.1515/cclm-2026-0844.
Katie Thoren, Lauren Campbell, Rosa Pello, María Ángeles Iglesias, Raúl Mateos Pablos, Cristina Agulló, et al (2026). Harmonizing M-protein analysis: an international, multi-center reproducibility study of mass spectrometry and evaluation of concordance to electrophoretic methods.. Clinical chemistry and laboratory medicine. https://doi.org/10.1515/cclm-2026-0844
Katie Thoren, Lauren Campbell, Rosa Pello, María Ángeles Iglesias, Raúl Mateos Pablos, Cristina Agulló, et al. Harmonizing M-protein analysis: an international, multi-center reproducibility study of mass spectrometry and evaluation of concordance to electrophoretic methods.[J]. Clinical chemistry and laboratory medicine, 2026 doi:10.1515/cclm-2026-0844.