新型可穿戴反馈装置对医学生心肺复苏训练的影响:一项三臂集群随机对照试验
The effect of a novel wearable feedback device on cardiopulmonary resuscitation training in medical students: a three-arm cluster randomized controlled trial.
The effect of a novel wearable feedback device on cardiopulmonary resuscitation training in medical students: a three-arm cluster randomized controlled trial.
尽管实时反馈设备已被证明能提高心肺复苏(CPR)培训中的技能掌握程度,但现有设备因成本高、皮肤刺激和功能单一等问题往往不尽理想。为克服这些局限,我们设计了一种新型可穿戴反馈设备,其准确性已在我们前期研究中得到验证。此外,我们构想了一种将实时反馈与刻意练习相结合的混合培训模式。鉴于目前缺乏该方法的有效性数据,本研究旨在评估该设备的实际培训表现,并探讨该混合培训模式能否进一步优化复苏培训效果。研究招募了重庆医科大学的学生,并随机分为三组:传统培训组(A组)、实时反馈培训组(B组)和实时反馈加刻意练习培训组(C组)。所有参与者均需在CPR模型上实施2分钟胸外按压。主要结局指标为正确按压深度(50-60 mm)的比例和 adequate 按压频率(100-120次/分)的比例。次要结局指标为胸外按压深度(mm)、胸外按压频率(次/分)以及正确按压回弹的比例。共618名参与者被随机分配至A组(n=210)、B组(n=204)和C组(n=204)。B组正确按压深度比例高于A组(均值[标准差],82.55[12.67]% vs. 71.54[22.28]%,均值差[95%置信区间],-11.0[-14.5至-7.5]%,P<0.001),低于C组(90.01[8.32]%,均值差,7.5[5.4至9.6]%,P<0.001)。B组 adequate 按压频率比例高于A组(均值[标准差],81.61[14.59]% vs. 78.15[13.60]%,均值差[95%置信区间],-3.5[-6.3至-0.7]%,P=0.014),与C组(83.12[11.93]%,均值差[95%置信区间],1.5[-1.1至4.1]%)无显著差异。B组按压深度深于A组(均值[标准差],57.02[5.15] mm vs. 53.47(4.71)mm,P<0.001),但与C组(57.45[4.36] mm,P=0.367)无差异。三组间胸外按压频率或正确按压回弹比例均无显著差异(P>0.05)。总之,在医科大学的学术环境中,与传统培训相比,这种新型反馈设备改善了医学生CPR的按压深度和频率表现。此外,实时反馈与刻意练习相结合的混合培训方法在该人群中进一步提高了胸外按压深度方面的培训效果。然而,在提出广泛实施建议之前,这些发现需要在更广泛的人群和多样化的临床环境中进行独立验证。
While real-time feedback devices are proven to improve skill mastery in CPR training, current options are often suboptimal due to high costs, skin irritation, and single-functionality constraints. To overcome these limitations, we engineered a novel wearable feedback device, the accuracy of which was validated in our prior research. Furthermore, we conceptualized a hybrid training paradigm combining real-time feedback with deliberate practice. Given the current lack of efficacy data for this approach, this study aims to assess the device's real-world training performance and investigate whether this hybrid training approach can further optimize resuscitation training outcomes. Students from Chongqing Medical University were recruited and randomly assigned to three groups: traditional training (Group A), real-time feedback training (Group B), or real-time feedback plus deliberate practice training (Group C). All participants were required to perform 2 min of chest compressions on CPR manikins. The primary outcomes were the proportion of correct compression depth (50-60 mm) and adequate compression rate (100-120 min). The secondary outcomes were chest compression depth (mm), chest compression rate (min), and the proportion of correct compression release. A total of 618 participants were randomly assigned to Group A ( = 210), Group B ( = 204), and Group C ( = 204). The proportion of correct compression depth in Group B was higher than that in Group A (mean [SD], 82.55 [12.67] % vs. 71.54 [22.28] %, mean difference [95% CI], -11.0 [-14.5 to -7.5] %, < 0.001) and lower than that in Group C (90.01 [8.32] %, mean difference, 7.5 [5.4 to 9.6] %, < 0.001). The proportion of adequate compression rate in Group B was higher than that in Group A (mean [SD], 81.61 [14.59] % vs. 78.15 [13.60] %, mean difference [95% CI], -3.5 [-6.3 to -0.7] %, = 0.014) and showed no significant difference from that in Group C (83.12 [11.93] %, mean difference [95% CI], 1.5 [-1.1 to 4.1] %). The compression depth in Group B was deeper than that in Group A (mean [SD], 57.02 [5.15] mm vs. 53.47 (4.71) mm, < 0.001) but did not differ from that in Group C (57.45 [4.36] mm, = 0.367). No significant differences were observed in chest compression rate or the proportion of correct compression release among the three groups ( > 0.05). In conclusion, within the academic setting of a medical university, this novel feedback device improved CPR performance in compression depth and rate among medical students compared to traditional training. Furthermore, the hybrid training method combining real-time feedback and deliberate practice further enhanced training efficacy regarding chest compression depth in this population. However, these findings require independent replication in broader populations and diverse clinical settings before widespread implementation recommendations can be made.
Enze Liu, Liyuan Gan, Wenwen Ma, Mo Chen, Chen Zhang, Xintao Huang, et al. The effect of a novel wearable feedback device on cardiopulmonary resuscitation training in medical students: a three-arm cluster randomized controlled trial.. Frontiers in public health. 2026; doi:10.3389/fpubh.2026.1883769.
Enze Liu, Liyuan Gan, Wenwen Ma, Mo Chen, Chen Zhang, Xintao Huang, et al (2026). The effect of a novel wearable feedback device on cardiopulmonary resuscitation training in medical students: a three-arm cluster randomized controlled trial.. Frontiers in public health. https://doi.org/10.3389/fpubh.2026.1883769
Enze Liu, Liyuan Gan, Wenwen Ma, Mo Chen, Chen Zhang, Xintao Huang, et al. The effect of a novel wearable feedback device on cardiopulmonary resuscitation training in medical students: a three-arm cluster randomized controlled trial.[J]. Frontiers in public health, 2026 doi:10.3389/fpubh.2026.1883769.
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出