Hemolysis and renal tubular injury after pulsed field ablation: A comparison between two commercially available platforms.
Hemolysis and renal tubular injury after pulsed field ablation: A comparison between two commercially available platforms.
脉冲电场消融(PFA)可引起血管内溶血,且不同平台之间的溶血程度存在差异。平台相关的溶血差异是否转化为肾小管损伤的差异尚未得到评估。旨在比较两种市售PFA平台在连续接受房颤消融患者中的溶血标志物、尿肾小管损伤生物标志物及短期肾功能。我们回顾性分析了131例连续患者,分别接受环形多电极阵列平台(PulseSelect;n=72)或五分支平台(Farapulse;n=59)治疗。于消融前一日及消融后一日测定血清结合珠蛋白、血红蛋白、乳酸脱氢酶、总胆红素和肌酐,以及尿N-乙酰-β-D-氨基葡萄糖苷酶(NAG)、β2-微球蛋白和α1-微球蛋白。比较两平台间的围术期变化。术后两组血清结合珠蛋白均显著下降,Farapulse组下降幅度更大(Δ结合珠蛋白:-59 [-71至-33] vs -36 [-51至-24] mg/dL;P < .001)。Farapulse组血清乳酸脱氢酶和总胆红素升高亦更明显(P = .002和P = .001)。尽管溶血反应更强,但尿N-乙酰-β-D-氨基葡萄糖苷酶、β2-微球蛋白和α1-微球蛋白的变化无显著差异(均P > .05),血清肌酐亦无差异(P = .948)。多变量回归分析显示,器械类型是Δ结合珠蛋白和Δ总胆红素的独立预测因子,但与任何肾小管生物标志物均无关联。在eGFR定义的慢性肾病患者中,平台相关的溶血差异持续存在,但尿肾小管生物标志物的变化仍相当。Farapulse在多个标志物上较PulseSelect与更强的血管内溶血相关,但在本研究的操作条件下,这些差异并未转化为可检测的肾小管损伤。
Pulsed field ablation (PFA) causes intravascular hemolysis, and the degree of hemolysis differs between available platforms. Whether platform-dependent differences in hemolysis translate into differences in renal tubular injury has not been evaluated. To compare hemolysis markers, urinary tubular injury biomarkers, and short-term renal function between 2 commercially available PFA platforms in consecutive patients undergoing AF ablation. We retrospectively analyzed 131 consecutive patients treated with a circular multielectrode array platform (PulseSelect; n = 72) or a pentaspline platform (Farapulse; n = 59). Serum haptoglobin, hemoglobin, lactate dehydrogenase, total bilirubin, and creatinine, and urinary N-acetyl-β-D-glucosaminidase (NAG), β2-microglobulin, and α1-microglobulin were measured the day before and the day after ablation. Periprocedural changes were compared between platforms. Serum haptoglobin decreased significantly after the procedure in both groups, with a greater decrease in the Farapulse group (ΔHaptoglobin: -59 [-71 to -33] vs -36 [-51 to -24] mg/dL; P < .001). Serum lactate dehydrogenase and total bilirubin also increased more with Farapulse (P = .002 and P = .001). Despite the greater hemolytic response, changes in urinary N-acetyl-β-D-glucosaminidase, β2-microglobulin, and α1-microglobulin did not differ significantly (all P > .05), nor did serum creatinine (P = .948). Device type was an independent predictor of ΔHaptoglobin and Δtotal bilirubin on multivariable regression but was not associated with any renal tubular biomarker. In patients with eGFR-defined chronic kidney disease, platform-dependent hemolysis persisted, but urinary tubular biomarker changes remained comparable. Farapulse was associated with greater intravascular hemolysis than PulseSelect across multiple markers, but these differences did not translate into detectable renal tubular injury under the procedural conditions of this study.
Keisuke Miyajima, Kazuku Ishihara, Takeru Yamada, Ryosuke Yabuzaki, Yusuke Oda, Kota Nanjo, et al. Hemolysis and renal tubular injury after pulsed field ablation: A comparison between two commercially available platforms.. Heart rhythm. 2026; doi:10.1016/j.hrthm.2026.07.043.
Keisuke Miyajima, Kazuku Ishihara, Takeru Yamada, Ryosuke Yabuzaki, Yusuke Oda, Kota Nanjo, et al (2026). Hemolysis and renal tubular injury after pulsed field ablation: A comparison between two commercially available platforms.. Heart rhythm. https://doi.org/10.1016/j.hrthm.2026.07.043
Keisuke Miyajima, Kazuku Ishihara, Takeru Yamada, Ryosuke Yabuzaki, Yusuke Oda, Kota Nanjo, et al. Hemolysis and renal tubular injury after pulsed field ablation: A comparison between two commercially available platforms.[J]. Heart rhythm, 2026 doi:10.1016/j.hrthm.2026.07.043.