深部脑刺激后脉冲发生器旋转导致充电失败:病例报告
Charging failure due to pulse generator rotation after deep brain stimulation: a case report.
Charging failure due to pulse generator rotation after deep brain stimulation: a case report.
一名54岁男性帕金森病患者因步态不稳及充电困难就诊,其困难在于为可充电式脑深部电刺激(DBS)脉冲发生器充电。胸部X线摄影显示该装置在皮下囊袋内发生水平旋转及前后倒置,此现象在既往影像中未见。手术翻修证实了装置方位异常,重新定位后恢复了充电效率。本病例提示,脉冲发生器旋转虽罕见但具有临床意义的硬件并发症,可在装置程控检测正常的情况下导致充电困难。这强调了在术后评估中考虑机械性原因并采用胸部影像学检查的重要性。
A 54-year-old man with Parkinson's disease presented with gait instability and difficulty charging his rechargeable deep brain stimulation (DBS) pulse generator. Chest radiography revealed horizontal rotation and anterior-to-posterior inversion of the device within the subcutaneous pocket, which was not seen in prior imaging. Surgical revision confirmed the device malorientation, and repositioning restored charging efficiency. This case highlights that pulse generator rotation, a rare but relevant hardware complication, can cause charging difficulties despite normal device interrogation. It underscores the importance of considering mechanical causes and using chest imaging in postoperative assessments.
Di Lu, Mingming Zhao, Feng Yin. Charging failure due to pulse generator rotation after deep brain stimulation: a case report.. Oxford medical case reports. 2026; doi:10.1093/omcr/omag137.
Di Lu, Mingming Zhao, Feng Yin (2026). Charging failure due to pulse generator rotation after deep brain stimulation: a case report.. Oxford medical case reports. https://doi.org/10.1093/omcr/omag137
Di Lu, Mingming Zhao, Feng Yin. Charging failure due to pulse generator rotation after deep brain stimulation: a case report.[J]. Oxford medical case reports, 2026 doi:10.1093/omcr/omag137.
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出