使用刺激电流诱导的非刺激电极电压在人工耳蜗植入术中视觉识别耳蜗外电极
Intraoperative Visual Identification of Extra-Cochlear Electrodes in Cochlear Implantation Using Stimulation Current-Induced Non-Stimulating Electrode Voltages.
Intraoperative Visual Identification of Extra-Cochlear Electrodes in Cochlear Implantation Using Stimulation Current-Induced Non-Stimulating Electrode Voltages.
耳蜗外电极(extra-cochlear electrodes, EEs)是人工耳蜗植入(cochlear implant, CI)手术中的常见并发症,可能导致言语感知结果下降。接触阻抗和电诱发复合动作电位等标准术中工具往往无法检测到这些位置异常的电极。本文介绍刺激电流诱导的非刺激电极电压(stimulation current-induced non-stimulating electrode voltages, SCINSEVs),即Advanced Bionics设备中的电场成像(electrical field imaging, EFI),作为一种更快速、更准确的术中识别耳蜗外电极的方法。在临床中采用该方法可减少对术中或术后影像学的依赖,并能在手术的关键时间窗内识别耳蜗外电极,从而可立即进行纠正。本前瞻性术中研究在单一机构纳入接受初次或翻修手术的成人人工耳蜗候选者或使用者。在电极插入过程中记录EFI测量值,术者被要求在完全插入前将随机选择的1至6个电极保持在耳蜗外。通过手术显微镜对耳蜗外电极位置数量进行视觉确认。记录在模拟术后条件下进行(耳蜗外电极被血液或筋膜包围)。三名对手术结果设盲的专家评审者审阅EFI数据以识别耳蜗外电极的数量。采用相关性分析、受试者工作特征曲线图以及灵敏度和特异度计算来分析评审者与术者之间的一致性。一般线性模型显示,术者未插入的耳蜗外电极数量(0至6个)和评审者(1至3名)对耳蜗外电极排序误差有显著影响,而手术类型(翻修或首次人工耳蜗植入)和所用材料(血液或筋膜)均无影响。未插入耳蜗外电极数量与基于EFI排序的耳蜗外电极之间观察到相关性(r = 0.63,p < 0.0001)。基于对SCINSEVs的视觉检查来识别耳蜗外电极的存在或不存在,总体灵敏度为95.9%,特异度为40.7%(在评审者与术者±2个耳蜗外电极一致性容差下分别为97.9%和65.7%)。SCINSEVs是一种快速、可靠的术中耳蜗外电极检测工具。其具有高灵敏度和中等特异度,可作为术后计算机断层扫描的非辐射替代方案,且结果可在手术过程中获得。使用该工具可通过立即手术纠正来改善人工耳蜗植入 outcomes。作者建议对相关临床工作人员进行培训,以记录和解读这些测量值,供临床实践使用。
Extra-cochlear electrodes (EEs) are a frequent complication in cochlear implant (CI) surgery, potentially leading to reduced speech-perception outcomes. Standard intraoperative tools such as contact impedances and electrically evoked compound action potentials often fail to detect these misplaced electrodes. Stimulation Current-Induced Non-Stimulating Electrode Voltages (SCINSEVs), or electrical field imaging (EFI) in the advanced bionics device, are presented here as a faster and more accurate intraoperative measure for identifying EE. Adopting this method clinically could reduce reliance on intraoperative or postoperative imaging and allow EE identification in the critical window of surgery, when immediate correction can occur. This prospective intraoperative study tested adult CI candidates or users undergoing new or revision surgery at a single institution. EFI measurements were recorded during electrode insertion, where the surgeon was asked to hold a randomly selected number of 1 to 6 electrodes outside of the cochlea before full insertion. Visual confirmation of the number of EE electrode positions was achieved with the surgical microscope. Recordings were taken simulating postoperative conditions (with EE surrounded by blood or fascia). Three expert reviewers blinded to surgical outcomes reviewed EFI data to identify the number of EEs. Agreement between reviewers and surgeons was analyzed using correlations, receiver operating characteristic plots, and sensitivity and specificity calculations. General linear models revealed significant effects of the number of EEs not inserted by the surgeon (0 to 6 EEs) and the reviewer (1 to 3) on the EE ranking error, and no effects of the surgery type (revision or first-time cochlear implantation), nor of the material used (blood or fascia). A correlation was observed between the number of not-inserted EEs and EEs ranked based on the EFIs (r = 0.63, p < 0.0001). For identification of the presence or absence of EEs based on visual inspection of SCINSEVs, an overall sensitivity of 95.9% and a specificity of 40.7% were achieved (97.9% and 65.7% with ±2 EE reviewer-surgeon agreement tolerance). SCINSEVs are a rapid, reliable tool for intraoperative EE detection. With high sensitivity and moderate specificity, they offer a non-radiative alternative to postoperative computerized tomography, with results available during the surgical procedure. Use of this tool may improve CI outcomes through immediate surgical correction. The authors recommend training of relevant clinical staff to record and interpret these measurements for use within clinical practice.
Daniele Borsetto, Charlotte Garcia, Marina Salorio-Corbetto, Susan Eitutis, Robert P Carlyon, James R Tysome, et al. Intraoperative Visual Identification of Extra-Cochlear Electrodes in Cochlear Implantation Using Stimulation Current-Induced Non-Stimulating Electrode Voltages.. Ear and hearing. 2026; doi:10.1097/AUD.0000000000001864.
Daniele Borsetto, Charlotte Garcia, Marina Salorio-Corbetto, Susan Eitutis, Robert P Carlyon, James R Tysome, et al (2026). Intraoperative Visual Identification of Extra-Cochlear Electrodes in Cochlear Implantation Using Stimulation Current-Induced Non-Stimulating Electrode Voltages.. Ear and hearing. https://doi.org/10.1097/AUD.0000000000001864
Daniele Borsetto, Charlotte Garcia, Marina Salorio-Corbetto, Susan Eitutis, Robert P Carlyon, James R Tysome, et al. Intraoperative Visual Identification of Extra-Cochlear Electrodes in Cochlear Implantation Using Stimulation Current-Induced Non-Stimulating Electrode Voltages.[J]. Ear and hearing, 2026 doi:10.1097/AUD.0000000000001864.
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出