基于光电容积描记法的指套式无袖带血压监测依据欧洲高血压学会推荐的临床验证

Clinical Validation of Photoplethysmography-Based Finger Cuffless Blood Pressure Monitoring Following European Society of Hypertension Recommendations.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

无袖带血压监测在临床高血压管理中仍具挑战性。尽管光电容积脉搏波描记法为无袖带血压测量提供了有前景的解决方案,但其临床验证仍然有限。本研究按照欧洲高血压学会的建议,评估了基于光电容积脉搏波描记法的指端无袖带血压监测的准确性和临床实用性。该两阶段研究共纳入1173名参与者。在第一阶段(n=768),一种新型基于光电容积脉搏波描记法的指端无袖带血压监测仪在7种欧洲高血压学会推荐的情景下,与有创血压、诊室血压及24小时动态血压监测进行了验证对比。在第二阶段(n=405名高血压患者),通过考察光电容积脉搏波描记法与动态血压监测所衍生的平均收缩压、收缩压目标范围内时间及变异性与高血压介导的靶器官损害和心血管并发症(包括动脉僵硬度、内皮功能障碍、冠心病和卒中)的关联,评估了其临床实用性。基于光电容积脉搏波描记法的指端无袖带血压监测仪符合所有验证标准,在包括有创、静态、清醒/睡眠、设备位置、治疗、运动及重新校准测试在内的7种情景下,收缩压和舒张压的平均差异均≤5 mmHg,标准差均≤8 mmHg。动态血压监测衍生的目标范围内时间以及光电容积脉搏波描记法衍生的平均收缩压和目标范围内时间均与内皮损伤和动脉僵硬度显著相关。值得注意的是,光电容积脉搏波描记法衍生的平均收缩压和目标范围内时间与冠状动脉疾病显示显著关联,而动态血压监测衍生的指标则无此关联。此外,光电容积脉搏波描记法衍生的收缩压变异性与内皮损伤呈显著正相关。我们的数据表明,这种新型基于光电容积脉搏波描记法的指端无袖带血压监测仪达到了欧洲高血压学会建议的临床应用标准,并为高血压介导的靶器官损害和心血管并发症的评估提供了额外的临床意义。

Abstract

Cuffless blood pressure (BP) monitoring remains challenging for clinical hypertension management. Although photoplethysmography offers promising solutions for cuffless BP measurement, its clinical validation is limited. This study evaluated the accuracy and clinical utility of photoplethysmography-based finger cuffless BP monitoring following European Society of Hypertension recommendations. The 2-phase study enrolled 1173 participants. In phase 1 (n=768), a novel photoplethysmography-based finger cuffless BP monitor was validated against invasive BP, office BP, and 24-hour ambulatory BP monitoring across 7 European Society of Hypertension-recommended scenarios. In phase 2 (n=405 hypertensive patients), the clinical utility of photoplethysmography- versus ambulatory BP monitoring-derived mean systolic BP (SBP), SBP time-in-target range, and variability was assessed by examining their associations with hypertension-mediated organ damage and cardiovascular complications, including arterial stiffness, endothelial dysfunction, coronary heart disease, and stroke. The photoplethysmography-based finger cuffless BP monitor fulfilled all validation criteria, with mean differences ≤5 mm Hg and standard deviations ≤8 mm Hg of both systolic and diastolic BP across 7 scenarios, including invasive, static, awake/asleep, device position, treatment, exercise, and recalibration test. Both ambulatory BP monitoring-derived time-in-target range and photoplethysmography-derived mean SBP and time-in-target range were significantly associated with endothelial injury and arterial stiffness. Notably, photoplethysmography-derived mean SBP and time-in-target range, but not ambulatory BP monitoring-derived metrics, showed a significant association with coronary artery disease. Furthermore, photoplethysmography-derived SBP variability was significantly and positively associated with endothelial injury. Our data demonstrated that the novel photoplethysmography-based finger cuffless BP monitor reached the criteria of clinical application following European Society of Hypertension recommendations and provided additional clinical implications for the evaluation of hypertension-mediated organ damage and cardiovascular complications.

如何引用

AMA

Yumin Qiu, Zhe Zhou, Jianning Zhang, Jiang He, Shiyue Xu, Xiaoyu Zhang, et al. Clinical Validation of Photoplethysmography-Based Finger Cuffless Blood Pressure Monitoring Following European Society of Hypertension Recommendations.. Journal of the American Heart Association. 2026; doi:10.1161/JAHA.126.050501.

APA

Yumin Qiu, Zhe Zhou, Jianning Zhang, Jiang He, Shiyue Xu, Xiaoyu Zhang, et al (2026). Clinical Validation of Photoplethysmography-Based Finger Cuffless Blood Pressure Monitoring Following European Society of Hypertension Recommendations.. Journal of the American Heart Association. https://doi.org/10.1161/JAHA.126.050501

GB/T 7714

Yumin Qiu, Zhe Zhou, Jianning Zhang, Jiang He, Shiyue Xu, Xiaoyu Zhang, et al. Clinical Validation of Photoplethysmography-Based Finger Cuffless Blood Pressure Monitoring Following European Society of Hypertension Recommendations.[J]. Journal of the American Heart Association, 2026 doi:10.1161/JAHA.126.050501.