扩张后的顺序很重要:近端优化技术(POT)减少了血管内超声评估的纵向支架变形

The Sequence of Post-Dilation Matters: Proximal Optimization Technique (POT) Reduces Longitudinal Stent Deformation Assessed by Intravascular Ultrasound.

获取 PDF 原文 在 PubMed增强版打开

摘要(中文译文)

纵向支架变形(LSD)是经皮冠状动脉介入治疗(PCI)的潜在严重并发症。尽管后扩张是LSD的已知危险因素,但扩张顺序等手术相关因素的影响仍知之甚少。本研究旨在识别与LSD相关的手术因素,重点关注后扩张技术。在529例接受血管内超声(IVUS)引导下PCI的患者(533条血管)中,于后扩张后测量支架长度。根据长度变化将患者分类:伸长(>2 mm或>名义支架长度的10%;n=71)、不变(±2 mm或<名义支架长度的10%;n=457)和缩短(>2 mm或>名义支架长度的10%;n=1,可能归因于测量误差,无统计学意义)。分析手术和病变特征。在单变量分析中,后扩张顺序(p<0.01)和球囊长度(p<0.05)在组间存在差异。在多变量logistic回归中,由远及近的扩张顺序(OR 3.13,95% CI 1.2-10.81)和较高的LVEF(OR 1.07,95% CI 1.02-1.12)与支架伸长独立相关。支架置于开口处具有保护作用(OR 0.45,95% CI 0.16-0.96)。在伸长组中,伸长幅度与病变长度、支架长度和扩张顺序呈正相关(均p<0.05)。后扩张顺序显著影响LSD。从支架近端段开始扩张——即近端优化技术(POT)——可能降低支架变形的风险。

Abstract

Longitudinal stent deformation (LSD) is a potentially serious complication of percutaneous coronary intervention (PCI). Although post-dilation is a known risk factor for LSD, the influence of procedural factors such as dilation sequence remains poorly understood. This study aimed to identify procedural factors associated with LSD, with a focus on post-dilation technique. In 529 patients (533 vessels) undergoing intravascular ultrasound (IVUS)-guided PCI, stent length was measured after post-dilation. Patients were categorized based on length change: elongation (> 2 mm or > 10% of nominal stent length; n = 71), unchanged (±2 mm or < 10% of nominal stent length; n = 457), and shortening (> 2 mm or > 10% of nominal stent length; n = 1, likely attributable to measurement error and not statistically significant). Procedural and lesion characteristics were analyzed. In univariate analysis, post-dilation sequence (p < 0.01) and balloon length (p < 0.05) differed between groups. In multivariate logistic regression, a distal-to-proximal dilation sequence (OR 3.13, 95% CI 1.2-10.81) and higher LVEF (OR 1.07, 95% CI 1.02-1.12) were independently associated with stent elongation. Stent placement at ostium was protective (OR 0.45, 95% CI 0.16-0.96). In the elongation group, positive correlations were observed between elongation magnitude and lesion length, stent length, and dilation sequence (all p < 0.05). The post-dilation sequence significantly influences LSD. Initiating dilation from the proximal stent segment-known as the proximal optimization technique (POT)-may reduce the risk of stent deformation.

如何引用

AMA

Dou Bo-Xu, Song Zi-Guang, Zhao Ji-Yi. The Sequence of Post-Dilation Matters: Proximal Optimization Technique (POT) Reduces Longitudinal Stent Deformation Assessed by Intravascular Ultrasound.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2026; doi:10.1002/ccd.70774.

APA

Dou Bo-Xu, Song Zi-Guang, Zhao Ji-Yi (2026). The Sequence of Post-Dilation Matters: Proximal Optimization Technique (POT) Reduces Longitudinal Stent Deformation Assessed by Intravascular Ultrasound.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. https://doi.org/10.1002/ccd.70774

GB/T 7714

Dou Bo-Xu, Song Zi-Guang, Zhao Ji-Yi. The Sequence of Post-Dilation Matters: Proximal Optimization Technique (POT) Reduces Longitudinal Stent Deformation Assessed by Intravascular Ultrasound.[J]. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions, 2026 doi:10.1002/ccd.70774.

在 PubMed增强版 阅读全文

翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出

获取 PDF 原文 在 PubMed增强版打开 PubMed 原文 DOI

Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions 的其他文献