Impact of field number and monitor units per segment on magnetic resonance-guided hypofractionated stereotactic radiotherapy for brain metastases: plan quality, deliverability, and robustness trade-offs.
Shan Ren, Jinrong He, Sheng Han, Nuo Xu, Wenyang Wang, Shipai Zhu, An Li, Shoupeng Liu, Jialu Lai, Xiaodong Peng, Renming Zhong
摘要(中文译文)
磁共振引导直线加速器(MR-Linacs)近年来被用于脑转移瘤(BMs)的放射治疗,包括大分割立体定向放射治疗(HSRT)。然而,在MR引导的自适应工作流程中制定HSRT计划的最佳策略仍有待确立。本研究旨在评估射野数量和每子野最小机器跳数(MU/segment)对计划质量的影响,以及MR引导HSRT治疗单发中型脑转移瘤患者时剂量计算的稳健性。本回顾性研究纳入20例接受HSRT治疗的单发中型脑转移瘤患者,处方剂量为30 Gy分3次照射。通过改变射野数量(9 vs 15)和最小MU/segment(15 vs 5),为每例患者系统性地生成四种调强放射治疗(IMRT)计划:9FL-IMRT(9个射野,15 MU/segment)、9FS-IMRT(9个射野,5 MU/segment)、15FL-IMRT(15个射野,15 MU/segment)和15FS-IMRT(15个射野,5 MU/segment)。基于使用基于结构的大体电子密度赋值优化的剂量分布,评估计划质量、治疗效率和照射准确性,并通过全局伽马通过率(GPR)在3%/2 mm、2%/2 mm和2%/1 mm标准下评估患者特异性质量保证(QA)。为评估对密度相关空间不确定性的稳健性,在模拟颅骨错位(随机平移偏移在±2 mm以内)后对每个计划进行重新优化。这些重新优化的计划在原始计算机断层扫描(CT)数据集上以基于体素的电子密度为参考标准进行重新计算。通过使用3%/2 mm、2%/2 mm、2%/1 mm和1%/1 mm标准的GPR分析比较基于大体密度和基于体素重新计算的剂量分布,量化稳健性。所有四种IMRT配置均提供了临床可接受的计划,靶区覆盖相似,适形指数、梯度指数或均匀性指数均无显著差异(均P>0.05)。尽管正常脑组织的一些剂量参数达到统计学显著性(总体P<0.001),但绝对差异较小(≤1 cm)且无临床意义。9FL-IMRT配置产生的MU最低(2,037±174),出束时间(BOT)最短(6.69±0.92 min),而15FL-IMRT配置仅需极小的BOT增加(7.59±1.18 min),并保持较少的子野数(33±7)。所有技术的照射准确性均持续较高,在3%/2 mm标准下平均GPR值≥95%,技术间无显著差异(P=0.67)。然而,在模拟颅骨位移条件下,更严格标准下出现了显著的稳健性差异,15FL-IMRT在2%/1 mm(96.02%±2.93%,总体P=0.004)和1%/1 mm(91.87%±3.55%,总体P<0.001)标准下GPR始终最高,显著优于9FL-IMRT和9FS-IMRT,表明其对密度相关空间不确定性的耐受性增强。15FL-IMRT配置在计划质量、可实施性和剂量计算稳健性之间提供了最佳平衡,支持其作为单发中型脑转移瘤患者MR引导HSRT首选计划方法的采用。
Abstract
Magnetic resonance-guided linear accelerators (MR-Linacs) have recently been introduced for radiotherapy of brain metastases (BMs), including hypofractionated stereotactic radiotherapy (HSRT). However, optimal strategies for planning HSRT within MR-guided adaptive workflows remain to be established. This study aimed to evaluate the influence of field number and minimum monitor unit per segment (MU/segment) on plan quality, and robustness of dose calculations in MR-guided HSRT for patients with solitary medium-sized BMs. This retrospective study included 20 patients who underwent HSRT for solitary medium-sized BMs, receiving a prescription dose of 30 Gy in three fractions. Four intensity-modulated radiotherapy (IMRT) plans were systematically generated for each patient by varying the number of fields (9 15) and minimum MU/segment (15 5): 9FL-IMRT (9 fields, 15 MU/segment), 9FS-IMRT (9 fields, 5 MU/segment), 15FL-IMRT (15 fields, 15 MU/segment), and 15FS-IMRT (15 fields, 5 MU/segment). Plan quality, treatment efficiency, and delivery accuracy were assessed based on dose distributions optimized using structure-based bulk electron density assignment, with patient-specific quality assurance (QA) evaluated via global gamma passing rate (GPR) using 3%/2 mm, 2%/2 mm, and 2%/1 mm criteria. To assess robustness against density-related spatial uncertainties, each plan was reoptimized after simulating skull misalignment with random translational offsets within ±2 mm. These reoptimized plans were recalculated on original computed tomography (CT) datasets using voxel-based electron density as the reference standard. Robustness was quantified by comparing dose distributions between bulk density-based and voxel-based recalculations through GPR analysis using 3%/2 mm, 2%/2 mm, 2%/1 mm, and 1%/1 mm criteria. All four IMRT configurations provided clinically acceptable plans with similar target coverage and no significant differences in conformity, gradient, or homogeneity indices (all P>0.05). Although some dose parameters for normal brain tissue reached statistical significance (overall P<0.001), absolute differences were minor (≤1 cm) and clinically irrelevant. The 9FL-IMRT configuration yielded the lowest MU (2,037±174) and shortest beam-on time (BOT; 6.69±0.92 min), while the 15FL-IMRT configuration required only a minimal increase in BOT (7.59±1.18 min) and maintained fewer segments (33±7). Delivery accuracy was consistently high across all techniques, with mean GPR values ≥95% at 3%/2 mm, showing no significant inter-technique differences (P=0.67). However, under conditions of simulated skull displacement, notable robustness differences emerged under stricter criteria, with 15FL-IMRT consistently showing the highest GPR at 2%/1 mm (96.02%±2.93%, overall P=0.004) and 1%/1 mm (91.87%±3.55%, overall P<0.001), significantly outperforming 9FL-IMRT and 9FS-IMRT, indicating enhanced tolerance to density-related spatial uncertainties. The 15FL-IMRT configuration provided an optimal balance among plan quality, deliverability, and robustness of dose calculation, supporting its adoption as the preferred planning approach for MR-guided HSRT in patients with solitary medium-sized BMs.
如何引用
AMAShan Ren, Jinrong He, Sheng Han, Nuo Xu, Wenyang Wang, Shipai Zhu, et al. Impact of field number and monitor units per segment on magnetic resonance-guided hypofractionated stereotactic radiotherapy for brain metastases: plan quality, deliverability, and robustness trade-offs.. Quantitative imaging in medicine and surgery. 2026; doi:10.21037/qims-2025-1-2798.
APAShan Ren, Jinrong He, Sheng Han, Nuo Xu, Wenyang Wang, Shipai Zhu, et al (2026). Impact of field number and monitor units per segment on magnetic resonance-guided hypofractionated stereotactic radiotherapy for brain metastases: plan quality, deliverability, and robustness trade-offs.. Quantitative imaging in medicine and surgery. https://doi.org/10.21037/qims-2025-1-2798
GB/T 7714Shan Ren, Jinrong He, Sheng Han, Nuo Xu, Wenyang Wang, Shipai Zhu, et al. Impact of field number and monitor units per segment on magnetic resonance-guided hypofractionated stereotactic radiotherapy for brain metastases: plan quality, deliverability, and robustness trade-offs.[J]. Quantitative imaging in medicine and surgery, 2026 doi:10.21037/qims-2025-1-2798.
在 PubMed增强版 阅读全文
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出
Quantitative imaging in medicine and surgery 的其他文献