优化经鼻温度调节:重症患者单鼻孔冷却装置的有效性和安全性
Optimizing Transnasal Temperature Modulation: Efficacy and Safety of a Single-Nostril Cooling Device in Critically Ill Patients.
Optimizing Transnasal Temperature Modulation: Efficacy and Safety of a Single-Nostril Cooling Device in Critically Ill Patients.
对脑血管损伤危重患者进行有效的体温管理至关重要。经鼻体温调节装置(tnTMD)在实现正常体温方面已显示出应用前景,但其双鼻孔配置的局限性影响了其疗效。本研究评估单鼻孔配置的有效性和安全性,并确定实现有效降温的最佳流速。这是一项前瞻性、多中心、开放标签研究,采用tnTMD进行降温以实现正常体温。30例患有缺血性或出血性卒中、癫痫发作或代谢性脑病的成年患者接受了tnTMD降温治疗。10例患者采用双鼻孔配置以60 L/min治疗8小时,另10例患者以相同流速治疗24小时。10例患者采用单鼻孔配置治疗24小时,流速逐步递增(30、40和50 L/min)。记录核心体温、寒战和不良事件。统计分析比较了单鼻孔队列中各流速下的体温下降幅度和安全性结局,还比较了tnTMD双鼻孔治疗与单鼻孔治疗的疗效数据。共有10例患者接受了单鼻孔降温。40 L/min流速在4小时内实现正常体温最为稳定,与30 L/min相比,24小时内的体温负荷降低(= 0.03)。40和50 L/min的降温疗效相似,但40 L/min的方差更小。3例出现寒战,但未观察到严重寒战病例。未出现严重不良效应。单鼻孔装置显示出与双鼻孔方法相当的疗效,且实用性更佳。tnTMD通过新型单鼻孔设计有效实现正常体温,最佳流速为40 L/min。鉴于其使用便捷且安全性良好,未来研究应探索其在非插管患者及更长时间降温干预中的适用性。
Effective temperature management in critically ill patients with cerebrovascular injuries is vital. A transnasal temperature modulation device (tnTMD) has shown promise in achieving normothermia, but challenges with its dual-nostril configuration limited its efficacy. This study evaluates the efficacy and safety of a single-nostril configuration and identifies the optimal flow rate for effective cooling. This was a prospective, multi-center, open-label study of temperature reduction to achieve normothermia utilizing a tnTMD. Thirty adult patients with ischemic or hemorrhagic stroke, seizure, or metabolic encephalopathy underwent cooling with the tnTMD. Ten patients were treated with the dual-nostril configuration for 8 hours at 60 L/min, and another 10 patients for 24 hours at the same flow rate. Ten patients were treated with the single-nostril configuration for 24 hours at incrementally escalating flow rates (30, 40, and 50 L/min). Core temperature, shivering, and adverse events were reported. The statistical analysis compared reductions in temperature and safety outcomes across flow rates in the single-nostril cohort. The analysis also compares efficacy data from the dual-nostril treatment with those from single-nostril treatments of the tnTMD. A total of 10 patients underwent single-nostril cooling. The 40 L/min flow rate achieved the most consistent normothermia within 4 hours, with reduced temperature burden over 24 hours compared with 30 L/min ( = 0.03). Cooling efficacy at 40 and 50 L/min was similar, but 40 L/min showed lower variance. Shivering occurred in three cases; however, no severe cases of shivering were observed. There were no severe adverse effects. The single-nostril device demonstrated efficacy comparable to that of the dual-nostril approach, with improved practicality. The tnTMD effectively achieved normothermia with a novel single-nostril design with an optimal flow rate of 40 L/min. Given its ease of use and benign safety profile, future research should explore its applicability in non-intubated patients and in longer-duration cooling intervention.
Melissa Wagner, Kenneth Dalton, Dave Peprah, Jason J Chang, H Alex Choi, Neeraj Badjatia. Optimizing Transnasal Temperature Modulation: Efficacy and Safety of a Single-Nostril Cooling Device in Critically Ill Patients.. Therapeutic hypothermia and temperature management. 2026; doi:10.1177/21537658261468190.
Melissa Wagner, Kenneth Dalton, Dave Peprah, Jason J Chang, H Alex Choi, Neeraj Badjatia (2026). Optimizing Transnasal Temperature Modulation: Efficacy and Safety of a Single-Nostril Cooling Device in Critically Ill Patients.. Therapeutic hypothermia and temperature management. https://doi.org/10.1177/21537658261468190
Melissa Wagner, Kenneth Dalton, Dave Peprah, Jason J Chang, H Alex Choi, Neeraj Badjatia. Optimizing Transnasal Temperature Modulation: Efficacy and Safety of a Single-Nostril Cooling Device in Critically Ill Patients.[J]. Therapeutic hypothermia and temperature management, 2026 doi:10.1177/21537658261468190.
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出