双能CT中不同碘浓度和流速下虚拟平扫成像及相关指标的验证:一项体模研究
Validation of virtual non-contrast imaging and related metrics with variable iodine concentrations and flow rates in dual-energy CT: a phantom study.
Validation of virtual non-contrast imaging and related metrics with variable iodine concentrations and flow rates in dual-energy CT: a phantom study.
验证虚拟平扫(VNC)成像准确性及双能CT(DECT)相关指标(CT值、CM值及碘密度值)在不同碘浓度和流速下的表现,采用体模模型进行研究。将碘溶液(3.7–18.5 mg I/mL)以五种流速(120–500 mL/min)循环通过体模。每种条件下使用SOMATOM go.Top进行DECT扫描三次。每次扫描在十张中心层面的感兴趣区(ROI)上测量四项指标(每种条件n = 3;N = 75)。采用双因素方差分析及Tukey HSD检验(α = 0.05)。碘浓度是所有指标的主要决定因素(p < .001)。VNC值呈现一致的负向偏差,且随浓度升高而增大,各条件间变异性较大。碘密度在浓度与流速之间观察到显著交互作用(p = .019),表明血流动力学影响物质分解。CT值、CM值及碘密度值呈现稳定、可预测的响应。VNC准确性在高碘浓度下受损,流速相关扰动通过物质分解误差引入额外的不稳定性。在解读VNC成像时应考虑血流动力学因素,尤其是在动脉期CTA中。
To validate virtual non-contrast (VNC) imaging accuracy and related dual-energy CT (DECT) metrics (CT, CM, and iodine density values) under varied iodine concentrations and flow rates using a phantom model. Iodine solutions (3.7-18.5 mg I/mL) were circulated through a phantom at five flow rates (120-500 mL/min). DECT scanning (SOMATOM go.Top) was performed three times per condition. Four metrics were measured from ROIs on ten central slices per scan (n = 3 per condition ; N = 75). Two-way ANOVA and Tukey's HSD test were applied (α = 0.05). Iodine concentration was the predominant determinant for all metrics (p < .001). VNC values showed a consistent negative deviation that increased with concentration and high variability across conditions. A significant interaction between concentration and flow rate was observed for iodine density (p = .019), indicating that flow dynamics affect material decomposition. CT, CM, and iodine density values showed stable, predictable responses. VNC accuracy is compromised at high iodine concentrations, with flow-related disturbances introducing additional instability via material decomposition errors. Flow dynamics should be considered when interpreting VNC imaging, particularly in arterial-phase CTA.
Masami Tashiro, Toshinori Shoji, Masaki Saito, Yuta Kaneko, Hideaki Takasumi. Validation of virtual non-contrast imaging and related metrics with variable iodine concentrations and flow rates in dual-energy CT: a phantom study.. Fukushima journal of medical science. 2026; doi:10.5387/fms.25-00045.
Masami Tashiro, Toshinori Shoji, Masaki Saito, Yuta Kaneko, Hideaki Takasumi (2026). Validation of virtual non-contrast imaging and related metrics with variable iodine concentrations and flow rates in dual-energy CT: a phantom study.. Fukushima journal of medical science. https://doi.org/10.5387/fms.25-00045
Masami Tashiro, Toshinori Shoji, Masaki Saito, Yuta Kaneko, Hideaki Takasumi. Validation of virtual non-contrast imaging and related metrics with variable iodine concentrations and flow rates in dual-energy CT: a phantom study.[J]. Fukushima journal of medical science, 2026 doi:10.5387/fms.25-00045.