光子计数CT定量X射线成像的潜力:基于有效原子序数和物理密度的新分析
Potential of quantitative X-ray imaging from photon-counting CT: A novel analysis based on effective atomic number and physical density.
Potential of quantitative X-ray imaging from photon-counting CT: A novel analysis based on effective atomic number and physical density.
传统肺部病变的CT评估主要基于形态学特征。CT值代表相对X射线衰减,并不直接反映组织成分和/或物理密度。利用光子计数CT(PC-CT)进行定量成像,有望提供定量参数,如有效原子序数(Z)和有效物理密度(ρ)。本研究旨在开发并验证一种算法的潜力,该算法可直接从临床PC-CT系统采集的虚拟单能图像(VMI)中生成Z图像和ρ图像。ρ图像是通过将质量衰减系数(μ/ρ)数据库拟合至测得的线性衰减系数(μ)而生成的,后者由70和100 keV的两个VMI获得。作为有效材料,Z信息也被确定并反馈至ρ的计算中。扫描了自制的低密度体模以对ρ进行定量评估。此外,为证明本流程适用于实际临床成像,分析了具有代表性的患者胸部PC-CT图像。在低密度体模实验中,CT值不一定与Z值相关,但与ρ值显示出非常好的相关性。使用本流程计算的ρ值与ρ的正确值相关性良好。ρ计算的相对均方根误差为5.5%。此外,我们发现,在肺部诊断中,与来自Z的对比信息相比,来自ρ的图像对比信息更容易识别和区分病变。所提出的流程直接从PC-CT数据生成Z图像和ρ图像,使得能够从元素组成和物理密度方面对CT值进行定量解读。本研究证明了生成Z图像和ρ图像的可行性。
Conventional CT assessment of lung lesions is based mainly on morphological features. CT values represent relative X-ray attenuation that does not directly reflect tissue composition and/or physical density. Quantitative imaging using photon-counting CT (PC-CT) has the potential to provide quantitative parameters, such as effective atomic number (Z) and effective physical density (ρ). The purpose is to develop and demonstrate the potential of an algorithm that generates both the Z and ρ images, derived directly from virtual monoenergetic images (VMIs) acquired with a clinical PC-CT system. The ρ image was generated by fitting a database of mass attenuation coefficients (μ/ρ) to the measured linear attenuation coefficients (μ), which were obtained from two VMIs at 70 and 100 keV. As an effective material, Z information was also determined and fed back into the ρ calculation. An in-house low-density phantom was scanned for quantitative evaluation of ρ. In addition, to demonstrate the applicability of our procedure to actual clinical imaging, representative chest PC-CT images from patients were analyzed. In experiments using low-density phantoms, the CT values did not necessarily correlate with Z values, but they showed a very good correlation with the ρ values. The ρ values calculated using our procedure correlated well with the correct values of ρ. A relative root mean square error of ρ calculation was 5.5%. Furthermore, we found that in pulmonary diagnosis, image contrast information from ρ makes it easier to identify and distinguish lesions compared to that from Z. The proposed procedure directly generated Z and ρ images from PC-CT data and enabled quantitative interpretation of CT value in terms of elemental composition and physical density. In this study, the feasibility of generating Z and ρ images was demonstrated.
Takashi Asahara, Rina Nishigami, Natsumi Kimoto, Mana Mitani, Yusuke Morimitsu, Noriaki Akagi, et al. Potential of quantitative X-ray imaging from photon-counting CT: A novel analysis based on effective atomic number and physical density.. Journal of applied clinical medical physics. 2026; doi:10.1002/acm2.70673.
Takashi Asahara, Rina Nishigami, Natsumi Kimoto, Mana Mitani, Yusuke Morimitsu, Noriaki Akagi, et al (2026). Potential of quantitative X-ray imaging from photon-counting CT: A novel analysis based on effective atomic number and physical density.. Journal of applied clinical medical physics. https://doi.org/10.1002/acm2.70673
Takashi Asahara, Rina Nishigami, Natsumi Kimoto, Mana Mitani, Yusuke Morimitsu, Noriaki Akagi, et al. Potential of quantitative X-ray imaging from photon-counting CT: A novel analysis based on effective atomic number and physical density.[J]. Journal of applied clinical medical physics, 2026 doi:10.1002/acm2.70673.
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出