长轴与短轴视野PET扫描仪用于脑成像:一项体模研究
Long vs. short axial field-of-view PET scanners for brain imaging: a phantom study.
Long vs. short axial field-of-view PET scanners for brain imaging: a phantom study.
长轴视野(LAFOV)PET扫描仪相比传统短轴视野(SAFOV)系统具有显著更高的灵敏度,有望在脑成像中实现降低注射剂量、缩短采集时间和改善信噪比。然而,鉴于这些系统之间的几何差异,在将已有的SAFOV方案转化应用于LAFOV系统之前,评估这些差异是否影响图像质量和定量准确性十分重要。本研究有两个目标:(1)使用Hoffman脑模体比较SAFOV PET扫描仪(Siemens Biograph Vision)与LAFOV PET扫描仪(Siemens Biograph Vision Quadra)之间的图像质量(IQ)指标;(2)量化LAFOV系统在脑成像中提供的有效轴向灵敏度增益。共进行了两项模体实验:第一项评估IQ指标及外部本底活度的影响,第二项评估LAFOV扫描仪内不同轴向位置处的脑特异性灵敏度。总体而言,两台扫描仪之间的IQ指标高度可比,且受外部本底活度的影响极小。将模体置于LAFOV扫描仪的第一个探测器环位置时,相对于SAFOV系统可获得1.2–1.4倍的灵敏度提升;而将其置于轴向中心位置时,灵敏度增益可达3.4倍。这些结果表明,LAFOV PET系统在保持与SAFOV扫描仪定量一致性的同时提供了更高的灵敏度。这支持开发优化PET成像方案以减少注射剂量和辐射暴露,从而提高安全性,并使更多脆弱人群能够参与脑PET研究。
Long axial field-of-view (LAFOV) PET scanners offer substantially higher sensitivity compared to conventional short axial field-of-view (SAFOV) systems, potentially enabling reduced injected activity, shorter acquisition times, and improved signal-to-noise ratios in brain imaging. However, given the geometric differences between these systems, it is important to assess whether these differences impact image quality and quantitative accuracy before translating established SAFOV protocols to LAFOV systems. This study has two goals: (1) to compare image quality (IQ) metrics between a SAFOV PET scanner (Siemens Biograph Vision) and a LAFOV PET scanner (Siemens Biograph Vision Quadra) using a Hoffman brain phantom, and (2) to quantify the effective axial sensitivity gain provided by the LAFOV system for brain imaging. Two phantom experiments were performed: the first assessed IQ metrics and the influence of external background activity, while the second evaluated brain-specific sensitivity at different axial positions within the LAFOV scanner. Overall, IQ metrics were highly comparable between scanners and were minimally affected by external background activity. Positioning the phantom at the first detector ring of the LAFOV scanner yielded a 1.2-1.4 sensitivity increase relative to the SAFOV system, whereas positioning it at the axial centre resulted in a sensitivity gain of up to 3.4 . These findings indicate that LAFOV PET systems maintain quantitative consistency with SAFOV scanners while offering higher sensitivity. This supports the development of optimized PET imaging protocols to reduce injected activity and radiation exposure, thereby improving safety and enabling broader inclusion of vulnerable populations in brain PET studies.
Laura Providência, Valentina Turmacu, Philipp Mohr, Chris W J van der Weijden, Rudi A J O Dierckx, Adriaan A Lammertsma, et al. Long vs. short axial field-of-view PET scanners for brain imaging: a phantom study.. Frontiers in nuclear medicine. 2026; doi:10.3389/fnume.2026.1856651.
Laura Providência, Valentina Turmacu, Philipp Mohr, Chris W J van der Weijden, Rudi A J O Dierckx, Adriaan A Lammertsma, et al (2026). Long vs. short axial field-of-view PET scanners for brain imaging: a phantom study.. Frontiers in nuclear medicine. https://doi.org/10.3389/fnume.2026.1856651
Laura Providência, Valentina Turmacu, Philipp Mohr, Chris W J van der Weijden, Rudi A J O Dierckx, Adriaan A Lammertsma, et al. Long vs. short axial field-of-view PET scanners for brain imaging: a phantom study.[J]. Frontiers in nuclear medicine, 2026 doi:10.3389/fnume.2026.1856651.
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出