{"_":"Cross-modality validation of quantitative Y imaging using a custom hepatopulmonary shunt phantom and patient data.","sup":["90"]}
{"_":"Cross-modality validation of quantitative Y imaging using a custom hepatopulmonary shunt phantom and patient data.","sup":["90"]}
本研究分析了在钇-90填充的肝肺分流模型(HPSP)中,使用SPECT/CT、PET/CT和PET/MR三种模态进行剂量估算时的差异。同时评估了随机选取的接受钇-90微球治疗患者的剂量变异性。HPSP包括右肺1101 cm³、左肺863 cm³、健康肝实质690 cm³以及两个肝内肿瘤分别为202 cm³和9.5 cm³。将氯化钇-90活度用水稀释后注入模型。模型浸没于水中进行SPECT/CT、PET/CT和PET/MR采集。对模型各分区及接受钇-90玻璃微球治疗的肝肿瘤患者计算吸收剂量。PET/MR图像计算所得的肺分流分数(LSF)(12.0%)最接近实际LSF(11.03%),其次为PET/CT(12.2%)和SPECT/CT(13.4%)。实际剂量估算值与钇-90 PET/MR之间的偏差为:大肿瘤(202 cm³)-0.1%,小肿瘤(9.5 cm³)-10.5%,健康肝实质-7.5%,肺6.1%。对于钇-90 PET/CT,偏差为大肿瘤4.8%,小肿瘤-7.5%,健康肝实质-15.5%,肺7.6%。相比之下,钇-90 SPECT/CT模态观察到显著差异,偏差为大肿瘤-6.3%,小肿瘤-65.3%,健康肝实质-10.7%,肺18.2%。在患者研究中,锝-99m MAA SPECT/CT与钇-90 PET/MR图像的剂量相当,而钇-90 SPECT/CT产生的剂量估算值显著偏低。研究结果表明,钇-90 SPECT/CT在吸收剂量估算中主要对小肿瘤产生显著不准确性。除SPECT/CT(18%)外,所有模态对LSF和肺剂量的高估均小于10%。结论认为,采用具有稳健散射校正的最优成像技术可减轻剂量低估。钇-90 PET/CT和钇-90 PET/MR是实现可靠剂量学和精确剂量-效应评估的首选方法,尤其是因为它们与锝-99m MAA SPECT/CT模拟具有良好的相关性。
In this study the discrepancies in dose estimations from a Y-filled hepatopulmonary shunt phantom (HPSP) were analyzed using SPECT/CT, PET/CT, and PET/MR modalities. The dose variability was also evaluated among randomly selected individuals undergoing Y-microsphere therapy. The HPSP comprised a right lung of 1,101 cm³, a left lung of 863 cm³, a healthy liver parenchyma of 690 cm³, and two intrahepatic tumours of 202 cm³ and 9.5 cm³. Y-chloride activity was diluted in water and injected into the phantom. SPECT/CT, PET/CT, and PET/MR were acquired with the phantom immersed in water. Absorbed dose was calculated for phantom compartments and for patients with liver tumours who were treated with Y-glass microspheres. The calculated lung shunt fraction (LSF) (12.0%) from the PET/MR images was closest to the actual LSF (11.03%), followed by PET/CT with 12.2% and SPECT/CT with 13.4% LSF. The deviation between the actual dose estimates and Y-PET/MR was - 0.1% for the large tumour (202 cm³), -10.5% for the small tumour (9.5 cm³), -7.5% for the healthy liver parenchyma, and 6.1% for the lung. For Y-PET/CT, the deviation was 4.8% for the large tumour, -7.5% for the small tumour, -15.5% for the healthy liver parenchyma, and 7.6% for the lung. In contrast, a significant disparity was observed in the Y-SPECT/CT modality, with deviations of -6.3% for the large tumour, -65.3% for the small tumour, -10.7% for the healthy liver parenchyma, and 18.2% for the lung. In the patient study, the doses from Tc-MAA SPECT/CT and Y-PET/MR images were comparable, whereas Y-SPECT/CT yielded significantly lower dose estimates. The findings obtained demonstrate that Y-SPECT/CT generates significant inaccuracies in absorbed dose estimates mainly for small tumours. All modalities overestimated LSF and lung dose by less than 10% except SPECT/CT (18%). It is concluded that implementing an optimal imaging technique with robust scatter correction would mitigate the dose underestimate. Y-PET/CT and Y-PET/MR are the preferred methods for reliable dosimetry and precise dose-response assessment, especially owing to their robust correlation with Tc-MAA SPECT/CT simulation.
Nami Yeyin, Mohammad Abuqbeitah, Mustafa Demir, Sinem Akyol, Duygu Tunçman Kayaokay, F Fatih Kesmezacar, et al. {"_":"Cross-modality validation of quantitative Y imaging using a custom hepatopulmonary shunt phantom and patient data.","sup":["90"]}. Radiation and environmental biophysics. 2026; doi:10.1007/s00411-026-01236-5.
Nami Yeyin, Mohammad Abuqbeitah, Mustafa Demir, Sinem Akyol, Duygu Tunçman Kayaokay, F Fatih Kesmezacar, et al (2026). {"_":"Cross-modality validation of quantitative Y imaging using a custom hepatopulmonary shunt phantom and patient data.","sup":["90"]}. Radiation and environmental biophysics. https://doi.org/10.1007/s00411-026-01236-5
Nami Yeyin, Mohammad Abuqbeitah, Mustafa Demir, Sinem Akyol, Duygu Tunçman Kayaokay, F Fatih Kesmezacar, et al. {"_":"Cross-modality validation of quantitative Y imaging using a custom hepatopulmonary shunt phantom and patient data.","sup":["90"]}[J]. Radiation and environmental biophysics, 2026 doi:10.1007/s00411-026-01236-5.
翻译全文 · AI 文献问答 · PDF 原文下载 · 引用导出