振动网式与喷射式雾化器在儿科呼吸支持中的性能:多模式体外比较
Vibrating Mesh and Jet Nebulizer Performance in Pediatric Respiratory Support: A Multi-Modality In Vitro Comparison.
Vibrating Mesh and Jet Nebulizer Performance in Pediatric Respiratory Support: A Multi-Modality In Vitro Comparison.
本研究旨在评估有创通气与无创通气过程中雾化药物递送的体外表现,并在多种儿童通气模型中比较喷射雾化器(JN)与振动网孔雾化器(VMN)。研究比较了连续输出型喷射雾化器(Aquineb)与振动网孔雾化器(Aerogen Solo A-VMN)的药物递送性能。无创模型使用解剖结构正确的上气道模型和呼吸模拟器,模拟一名9个月大自主呼吸儿童。有创模型使用配备加热湿化器的机械通气机,连接儿童呼吸回路及气管导管。雾化器按制造商推荐流速以补充氧气驱动,并置于批准位置。通过面罩、机械通气、经鼻高流量治疗及吹气式递送方式评估绝对吸入剂量、递送速率和残留量。剂量采用分光光度法分析定量。在自主呼吸期间,A-VMN递送的剂量约为所评估JN的近两倍(P < 0.001),递送速率显著更快(P < 0.001),残留量更低(P < 0.0001)。在机械通气期间,A-VMN递送剂量增加超过3倍(P < 0.0001),递送更快(P < 0.0001),残留量减少(P < 0.001)。在经鼻高流量治疗期间,两种装置经鼻导管的递送均受气体流速影响,A-VMN始终递送更大剂量。在吹气式递送中,A-VMN递送的沙丁胺醇剂量显著更大。在所有测试的儿童呼吸支持方式中,与所评估的JN相比,VMN均表现出显著更优的剂量递送、更快的递送速率和更低的残留量。这些体外研究结果为基于证据的装置选择提供了重要性能数据,并值得开展临床研究,以确定在接受呼吸支持且需要雾化治疗的儿童人群中可能获得的治疗获益。
The aim of this study was to assess in vitro nebulized drug delivery during invasive and non-invasive ventilation, comparing jet nebulizers (JN) and vibrating mesh nebulizers (VMN) across various pediatric ventilation models. Drug delivery performance was compared between a continuous output JN (Aquineb) and VMN (Aerogen Solo A-VMN). The non-invasive model simulated a spontaneously breathing 9-month-old child using an anatomically correct upper airway model and breathing simulator. The invasive model used a mechanical ventilator with heated humidifier in a pediatric breathing circuit with an endotracheal tube. Nebulizers were driven with supplemental oxygen at manufacturer-recommended rates and positioned at approved locations. Absolute inhaled dose, delivery rate and residual volume were assessed using face mask, mechanical ventilation, high-flow nasal therapy and blow-by delivery methods. Dose was quantified using spectrophotometric analysis. During spontaneous breathing, A-VMN delivered almost double the dose of the evaluated JN ( < 0.001), with a significantly faster delivery rate ( < 0.001) and lower residual volume ( < 0.0001). During mechanical ventilation, A-VMN demonstrated a greater than 3-fold increase in delivered dose ( < 0.0001) and faster delivery ( < 0.0001), with reduced residual volume ( < 0.001). During high-flow nasal therapy, delivery via nasal cannula was affected by gas flow rate for both devices, with A-VMN consistently delivering greater doses. A-VMN delivered significantly greater salbutamol doses during blow-by delivery. VMN demonstrated significantly superior dose delivery, faster delivery rates and reduced residual volumes compared to the evaluated JN across all tested pediatric respiratory support modalities. These in vitro findings provide important performance data for evidence-based device selection and warrant clinical investigation to determine potential therapeutic benefits in pediatric populations requiring aerosol therapy during respiratory support.
Ronan MacLoughlin, Ann-Marie Crowe, Michael Scully, Brendan D Higgins. Vibrating Mesh and Jet Nebulizer Performance in Pediatric Respiratory Support: A Multi-Modality In Vitro Comparison.. Pharmaceutics. 2026; doi:10.3390/pharmaceutics18050575.
Ronan MacLoughlin, Ann-Marie Crowe, Michael Scully, Brendan D Higgins (2026). Vibrating Mesh and Jet Nebulizer Performance in Pediatric Respiratory Support: A Multi-Modality In Vitro Comparison.. Pharmaceutics. https://doi.org/10.3390/pharmaceutics18050575
Ronan MacLoughlin, Ann-Marie Crowe, Michael Scully, Brendan D Higgins. Vibrating Mesh and Jet Nebulizer Performance in Pediatric Respiratory Support: A Multi-Modality In Vitro Comparison.[J]. Pharmaceutics, 2026 doi:10.3390/pharmaceutics18050575.